Carestation
620/650/650c (A1)
Message
Priority
FiDes low
Low
FiEnf high
Medium
FiEnf low
Low
FiHal high
Medium
FiHal low
Low
FiIso high
Medium
FiIso low
Low
FiO2 high
Medium
FiO2 low
High
FiSev high
Medium
FiSev low
Low
Gas monitoring
Medium
not available
Gas controls
High
failure, O2 only
Gas controls
Medium
failure, O2 only
Increase low MV
Medium
limit
Inspiration
Medium
stopped
Internal failure.
High
System may shut
down.
Internal failure.
Medium
System may shut
down.
7-8
Cause
FiDes is less than alarm limit.
FiEnf is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes
FiEnf is less than alarm limit.
FiHal is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes
FiHal is less than alarm limit.
FiIso is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes
FiIso is less than alarm limit.
FiO2 is greater than high alarm limit.
FiO2 is less than low alarm limit.
FiSev is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes
FiSev is less than alarm limit.
Airway module hardware failure.
System communication failure.
Frame interface board and power
management board failure.
Low MV limit is off in SIMV VCV, SIMV
PCV, SIMV PCV-VG, CPAP + PSV, or
PSVPro
modes.
High airway pressure.
Power controller software failure.
Power controller software failure.
Action
Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
Set the alarm limits appropriately.
Check agent setting.
Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
Set the alarm limits appropriately.
Check agent setting.
Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
Set the alarm limits appropriately.
Check agent setting.
Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
Check the O2 setting. Recalibrate the
O2 cell and the airway module.
Check O2 setting. Check for leaks or
blockages in the patient circuit.
Set the alarm limits appropriately.
Check agent setting.
Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
Replace airway module. Then, turn
power off and back on between cases
to clear the alarm and receive module
data.
Contact an authorized service
representative.
Contact an authorized service
representative.
Increase low MV alarm limit to improve
patient disconnection detection.
Check system for blockages.
Contact an authorized service
representative.
Contact an authorized service
representative.