Flow And Pressure Calibration - GE Carestation 620 User's Reference Manual

Table of Contents

Advertisement

Carestation
620/650/650c (A1)

Flow and pressure calibration

5-4
Calibrate the flow sensors by opening the breathing circuit. On-
screen instructions are available through the Checkout menu or
System Setup - Calibration - Flow and Pressure.
1.
Pull down the canister lifter handle and remove the absorber
canister.
2.
Open the breathing system door. Push the release button to
unlock the breathing system.
3.
Lower the breathing system, and select Confirm.
Note
The breathing system should be completely disconnected from the
system (fully rotated down) to obtain an accurate calibration.
4.
Lift and reattach the breathing system.
5.
Close the breathing system door.
6.
Place the canister onto the canister base.
7.
Pull the canister lifter handle up to lock the canister into the
breathing system.
2076152-001