Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare

4 Functional Requirements

4.1 Electrical Requirements

Innova System require two specific power lines, two three phase for C1 cabinet (see
Illustration AC Interconnect without Fluoro UPS in
Connect the main power demand to the X–Ray generator – Jedi (C1 cabinet).
X–Ray generator is feed by a 126 kVA three phase power line.
For Generator Power Supply Features, see ref 60601.1 IEC.
NOTE:
In CE configuration, C2 cabinet is fed by C1 cabinet.
Table 3-13: ELECTRICAL REQUIREMENTS: System line voltage without Fluoro UPS
Peak Max Power
Frequency (Hz)
input (kVA)
150
50 or 60
Table 3-14: ELECTRICAL REQUIREMENTS: System line voltage with Fluoro UPS UL
Peak Max Power
Frequency (Hz)
input (kVA)
150
Table 3-15: ELECTRICAL REQUIREMENTS: System line voltage with Fluoro UPS CE
Peak Max Power
Frequency (Hz)
input (kVA)
150
50 or 60
NOTE:
PDB (CE) maximum rating is equal to 57 kVA.
PDB (UL) maximum rating is equal to 124 kVA.
Line impedance should be compliant with IEC 601.2.7 Refer to the table Max Line
Impedance for feeder line between Generator cabinet and Hospital in
Section 4,
When a Fluoro UPS is or will be installed, a Neutral line is mandatory. An IT or
Delta configuration power input requires the installation of an isolation transformer
(see transformer specification below)..
72
Mains Voltage
(V)
380 to 480 (3 ph,
± 10%)
Mains Voltage
(V)
60
480 (3 ph + N,
±10% )
Mains Voltage
(V)
380/400/415 (3 ph
+ N, ±10% )
Power and Grounding Requirements.
Chapter 6, Section 1,
Max. Line
Peak Current (A)
Impedance
(ohms)
0.09 to 0.12
228 to 180 for C1
Max. Line
Peak Current (A)
Impedance
(ohms)
0.09 to 0.12
180 for C1
Max. Line
Peak Current (A)
Impedance
(ohms)
0.09 to 0.12
228 for C1
Direction 5314528-2-1EN, Revision 1
Power Distribution).
Average Current
(A)
16 to 13
Average Current
(A)
13
Average Current
(A)
16
4 Functional Requirements
Mains Wire
4 wires (3 ph +
GND).
Mains Wire
5 wires (3 ph + N
+ GND)
Mains Wire
5 wires (3 ph + N
+ GND)
Chapter 6,