Electrical Power; Power Cord - GE Carestation 620 User's Reference Manual

Table of Contents

Advertisement

Electrical power

Power cord

WARNING
2076152-001
11 Specifications and theory of operation
Supply voltage
100-120, 220-240, or 120/220-240 Vac +/- 10% at 50 or
60 Hz
Average power
Less than 55 VA
consumption
Note: Measured through the power cord when operated at the ventilation
setting VCV, TV = 500 ml/min, RR = 12/min, I:E = 1:2 with nothing plugged
into auxiliary outlets and an airway module inserted. Measurements were
taken over a 15 minute time frame after system had been running for 10
minutes.
Voltage
100-120 Vac 220-240 Vac 100-120 Vac 220-240 Vac
Outlet circuit
(2) 2 A
breakers
(1) 3 A
Mains fused
10 A
System leakage
UL and CSA rated systems (U.S.A. and Canada): less
current limit - do
than 300 μamps for the system and all systems
not exceed
connected to electrical outlets.
IEC rated systems (not U.S.A. and Canada): less than
500 µamps for the system and all systems connected to
electrical outlets.
Note: Products connected to electrical outlets may
increase the leakage current above these limits.
Resistance to
Less than 0.2 Ω.
ground
Length
Voltage rating
Current capacity
Type
This equipment must be connected to a supply mains
with a protective earth to avoid the risk of electric shock.
(2) 1 A
None
(1) 2 A
8 A
2 A
5 meters
100 to 240 Vac
10 A for 220-240 Vac
15 A for 100-120 Vac
10 A for 120/220-240 Vac
Three conductor power supply cord (medical
grade where required).
None
2 A
11-9