Table of Contents

Advertisement

Voluson™ P8 / Voluson™ P6
Basic Service Manual
• Voluson™ P8 systems
• Voluson™ P6 systems
Document Number: 5723243
Revision: 6
0459

Advertisement

Table of Contents
loading

Summary of Contents for GE Voluson P8 BT16

  • Page 1 Voluson™ P8 / Voluson™ P6 Basic Service Manual • Voluson™ P8 systems • Voluson™ P6 systems Document Number: 5723243 Revision: 6 0459...
  • Page 3 • WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN WARNUNG ANWWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND (DE) NACHVOLLZIEHEN KÖNNEN. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
  • Page 4 ™P8 / V ™P6 OLUSON OLUSON DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL • ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, LA TRADUCCIÓN ES RESPONSABILIDAD DEL CLIENTE.
  • Page 5 ™P8 / V ™P6 OLUSON OLUSON DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL VAROITUS (FI) ΠΡΟΕΙ∆ΟΠΟΙΗΣΗ (EL) FIGYELMEZTETÉS (HU) • ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU. •...
  • Page 6 ™P8 / V ™P6 OLUSON OLUSON DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL VÝSTRAHA (CS) • DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK. • HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE. •...
  • Page 7 ™P8 / V ™P6 OLUSON OLUSON DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL ĮSPĖJIMAS (LT) • DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK. • HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE. • IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE ADVARSEL SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
  • Page 8 ™P8 / V ™P6 OLUSON OLUSON DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL ОСТОРОЖНО! (RU) UPOZORNENIE (SK) • DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA. • OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA ÖVERSÄTTNINGSTJÄNSTER.
  • Page 9 ™P8 / V ™P6 OLUSON OLUSON DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL (JA) (ZH-CN) (KO)
  • Page 10 GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 11 ™P8 / V ™P6 OLUSON OLUSON DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL LEGAL NOTES The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of Voluson™P8 / Voluson™P6. Voluson™P8 / Voluson™P6 may revise this publication from time to time without written notice.
  • Page 12 ™P8 / V ™P6 OLUSON OLUSON DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Revision History Revision Date Reason for change August 2015 Initial release February 2016 Update Replacement procedures and SW Images July 2016 Enhancement feedback May 2017 Update chapter 8 to add new figures August 2017 Add contents : Mode Transition , Software DVD storage, Software Upgrade Checklist May 2018...
  • Page 13: Table Of Contents

    ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL CHAPTER 1 Introduction Overview............1 - 1 Purpose of Chapter 1 .
  • Page 14 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL CHAPTER 2 Site Preparation Overview ............2 - 1 Purpose of Chapter 2 .
  • Page 15: Direction 5723243, Revision 6 Basic Service Manual

    ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL CHAPTER 3 Setup Instructions Overview............3 - 1 The Purpose of Chapter 3 .
  • Page 16 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Transducer Connection ........3 - 31 Printer Installation .
  • Page 17 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Software/Option Configuration ........3 - 71 Connectivity Setup .
  • Page 18 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL CHAPTER 4 Functional Checks Overview ............4 - 1 Purpose of Chapter 4 .
  • Page 19 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Mode Transition ........... 4 - 42 B Mode .
  • Page 20 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview ............5 - 1 Purpose of Chapter 5 .
  • Page 21 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL HDLive(3D) ..........5 - 17 HDLive 3D .
  • Page 22 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Peripherals............5 - 39 Printers .
  • Page 23 GE-Service ........
  • Page 24 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL CollectLogs ..........5 - 79 CheckLnkSpeed .
  • Page 25 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL CHAPTER 6 Service Adjustments Overview............6 - 1 Purpose of Chapter 6 .
  • Page 26 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview ............7 - 1 Purpose of Chapter Voluson™...
  • Page 27 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL DBM ......... . . 7 - 37 DPI .
  • Page 28 No Connection to the Network at All ......7 - 57 GE remote service connection ....... . 7 - 57 Tech Tips .
  • Page 29 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview............8 - 1 Purpose of Chapter 8 .
  • Page 30 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Disassembly Procedure ........8 - 26 Installation Procedure .
  • Page 31 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Replacement of Keyboard Assembly-TOP....... 8 - 46 Manpower .
  • Page 32 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Replacement of ASSY Bumper Base Rubber ......8 - 81 Manpower .
  • Page 33 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Manpower ..........8 - 100 Tools .
  • Page 34 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Removal Procedure ......... . . 8 - 115 Installation Procedure .
  • Page 35 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Installation Procedure ......... 8 - 132 DRFM (5497712) Installation Procedure .
  • Page 36 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Tools ............8 - 149 Preparations .
  • Page 37 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Replacement of CWD piggy ......... . 8 - 175 Manpower .
  • Page 38 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL CHAPTER 9 Renewal Parts Overview ............9 - 1 Purpose of Chapter 9 .
  • Page 39 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL CHAPTER 10 Care & Maintenance Overview............10 - 1 Periodic Maintenance Inspections .
  • Page 40 ™P8 / V ™P6 OLUSON OLUSO DIRECTION 5723243, R EVISION ASIC ERVICE ANUAL Meter Procedure ......... . 10 - 20 Enclosure Leakage Current Test .
  • Page 41: Overview

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the Voluson™ P8 / Voluson™ P6 and/or Voluson™...
  • Page 42: Typical Users Of The Basic Service Manual

    EVISION ASIC ERVICE ANUAL 1-1-3 Typical Users of the Basic Service Manual • GE Service Personnel (installation, maintenance, etc.). • Hospital’s Service Personnel • Contractors (Some parts of Chapter 2 - Pre-Installation) 1-1-4 Models Covered by this Manual Table 1-2 Voluson™...
  • Page 43: Purpose Of Operator Manual(S)

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 1-1-5 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the Voluson™ P8 / Voluson™...
  • Page 44: Standard Hazard Icons

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
  • Page 45: Product Icons

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-5 Product Icons Voluson™...
  • Page 46 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 1-5 Product Icons Voluson™ P8 / Voluson™ P6 LABEL/SYMBOL PURPOSE/MEANING LOCATION "Mains OFF" Indicates the power off position of the mains Rear of system at mains switch power switch.
  • Page 47: Safety Considerations

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment.
  • Page 48: Electrical Safety

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 1-3-4 Electrical Safety 1-3-4-1 Safe Practices To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The system is equipped with a three-conductor AC power cable.
  • Page 49 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 1-3-5 Auxiliary Devices Safety (cont?) Everybody who connects additional equipment to the signal input portion or signal output portion configures a medical system, and is therefore responsible that the system complies with the requirements of the system standard IEC 60601-1-1.
  • Page 50: Labels Locations

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 1-3-6 Labels Locations The Voluson™ P8 / Voluson™ P6 comes equipped with product labels and icons. These labels and icons represent pertinent information regarding the operation of the unit. Figure 1-1 Monitor caution label(on rear of the Voluson™...
  • Page 51: Main Label

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 1-3-6-1 Main Label The Main Label is located on the rear of the Voluson™ P8 / Voluson™ P6 system. Figure 1-2 Main Label (located on left rear of Voluson™ P8 / Voluson™ P6) Manufacturer Product name Weight...
  • Page 52: Tuv Mark

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 1-3-6-2 TUV mark TUV mark is located on the rear of the system. Figure 1-3 TUV mark 1-12 Section 1-3 - Safety Considerations...
  • Page 53: Multi Caution Label

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 1-3-6-3 Multi Caution Label The Multi caution label is located on the rear of the Voluson™ P8 / Voluson™ P6 System Figure 1-4 Multi Caution Label Symbol “Caution”...
  • Page 54: Dangerous Procedure Warnings

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 1-3-7 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT DANGER IN THIS EQUIPMENT.
  • Page 55: Battery Safety

    NOTE: Regulations vary for different countries. Dispose of a used battery in accordance with local regulations. USE ONLY BATTERIES APPROVED BY GE HEALTHCARE AS SUITABLE FOR USE WITH THE CAUTION Voluson™ P8 / Voluson™ P6 ULTRASOUND SYSTEM.
  • Page 56: Electromagnetic Compatibility (Emc)

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 1-4 Electromagnetic Compatibility (EMC) 1-4-1 What is EMC? Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment.
  • Page 57: Customer Assistance

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 1-5 Customer Assistance 1-5-1 Contact Information If this equipment does not work as indicated in this service manual or in the Basic User Manual, or if you require additional assistance, please contact the local distributor or appropriate support resource, as listed below.
  • Page 58: System Manufacturer

    Figure 1-5 Phone Number for Customer Assistance 1-5-2 System Manufacturer Table 1-6 System Manufacturer Manufacturer FAX Number GE Ultrasound Korea, Ltd. 9, Sunhwan-ro 214beon-gil, Jungwon-gu, +82 (0) 31-740-6436 Seongnam-si, Gyeonggi-do, Korea 1-18 Section 1-5 - Customer Assistance...
  • Page 59: Site Preparation

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Chapter 2 Site Preparation Section 2-1 Overview 2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a Voluson™ P8 / Voluson™...
  • Page 60: General Console Requirements

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 2-2 General Console Requirements 2-2-1 Environmental Requirements Table 2-2 Environmental Requirements Operating Operating Storage Heat Dissipation Storage Humidity Temperature Humidity Temperature 1200BTU/h...
  • Page 61: Voluson™ P8 / Voluson™ P6 Power Requirements

    The Voluson™ P8 / Voluson™ P6 is supplied with a main power plug, as standard. In the event that the unit arrives without a power plug, or with the wrong plug, contact your GE dealer. When necessary, the installation engineer will supply the appropriate power plug to meet the applicable local regulations.
  • Page 62: Emi Limitations

    RF gaskets touch metal. appropriate location. The interconnect cables are grounded and require ferrite beads and other shielding. Use GE- specified Cable length, material, and routing are all important; do not make any changes that do not meet harnesses and peripherals.
  • Page 63: Probe Environmental Requirements

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2-2-4 Probe Environmental Requirements Operation: Ambient temperature 18° to 30° C Storage: -10° to 50° C NOTE: Temperature in degrees C. Conversion to degrees F = °C (9/5) + 32. SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF NOTICE -10 TO + 50...
  • Page 64: Facility Needs

    The products involved (and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
  • Page 65: Mandatory Site Requirements

    NOTE: All relevant preliminary network port installations at the prepared site must be performed by authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to perform servicing of the equipment. 2-3-3 Site Recommendations The following are (optional) site recommendations.
  • Page 66: Recommended Ultrasound Room Layout

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2-3-3-1 Recommended Ultrasound Room Layout Figure 2-1 below shows a floor plan illustrating the recommended layout of the Ultrasound Room and depicting the minimal room layout requirements.
  • Page 67: Networking Setup Requirements

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2-3-4 Networking Setup Requirements 2-3-4-1 Stand-alone Unit (without Network Connection) None 2-3-4-2 Unit Connected to Hospital’s Network Supported networks: • Ethernet •...
  • Page 68 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 2-10 Section 2-3 - Facility Needs...
  • Page 69: Setup Instructions

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Chapter 3 Setup Instructions Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information needed to setup the Voluson™ P8 / Voluson™ P6. Included are procedures to receive, unpack and configure the equipment.
  • Page 70: Set Up Reminders

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 3-2 Set Up Reminders 3-2-1 Average Installation Time Once the site has been prepared, the average installation time required is shown in Table 3-2 below.
  • Page 71: Installation Warnings

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-2-2 Installation Warnings 1.) Since the Voluson™ P8 / Voluson™ P6 weighs approximately 70 kg (154.3 lbs.) without peripherals, two people are required to unpack it. 2.) There are no operator serviceable components.
  • Page 72: Brake Pedal Operation

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-2-2-3 Brake Pedal Operation WARNING WARNING REMEMBER: If the front wheel brakes are engaged for transportation, release brake pedals (brakes on front wheels under the foot rest) to disengage the lock. !! WARNING: Section 3-2 - Set Up Reminders...
  • Page 73: Safety Reminders

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-2-3 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE DANGER AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT !! DANGER: TOUCH THE UNIT! Two people should unpack the unit because of its weight.
  • Page 74: Receiving And Unpacking The Equipment

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment Please read this section carefully before unpacking the Voluson™ P8 / Voluson™ P6 and its CAUTION (optional) peripherals.
  • Page 75 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont?) The envelope with delivery address, packing list and invoice is located on the front panel of the crate. Check whether delivery is complete (according to packing list) and check visual damage! Figure 3-4 envelope at front panel of the crate...
  • Page 76 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 3-5 Unpacking Procedure Step Task Remove the top cover and top plate. Remove the three Plastic Joints from the outer sleeve.. Remove the outer sleeve and the monitor pad.
  • Page 77 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 3-5 Unpacking Procedure Step Task Cut the tape to unfold the package and remove the antistatic vinyl cover Disengage the brakes, grasp the system at the handle, then slowly move unit down the ramp. Rear panel is used for ramp Note: Packing crate and material should be stored for future use.
  • Page 78: Preparing For Set Up

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 3-4 Preparing for Set Up 3-4-1 Verify Customer Order 1.) After unpacking the equipment, it is important to verify that all items ordered by the customer have been received.
  • Page 79 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL this interference. For this reason, it is imperative that all covers and hardware are installed and secured before the unit is put into operation. Ensure that the system is protected from electromagnetic interference (EMI), as follows: •...
  • Page 80: Connection Of Auxiliary Devices

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 3-5 Connection of Auxiliary Devices NOTE: Normally auxiliary devices and peripherals come pre-installed with the system. Table 3-7 below outlines hardware installation procedures described in the sub-sections. Table 3-7 Connection Procedures Sub-section...
  • Page 81: Connecting The Lcd Monitor

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-5-1 Connecting the LCD Monitor NOTE: The LCD monitor comes pre-installed with the system. Figure 3-5 Connection Scheme - LCD Monitor Chapter 3 - Setup Instructions 3-13...
  • Page 82: Connecting The Black & White Printer

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-5-2 Connecting the Black & White Printer 1.) Power OFF/Shutdown the system as described in: Section 3-6-3 on page 3-29. 2.) Connect the Black &...
  • Page 83: Connection Scheme: B&W Printer

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-5-2-1 Connection Scheme: B&W Printer Figure 3-6 B&W Printer connection Chapter 3 - Setup Instructions 3-15...
  • Page 84: Connecting The Color Printer

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-5-3 Connecting the Color Printer 1.) Power OFF/Shutdown the system as described in: Section 3-6-3 on page 3-29. 2.) Connect the Color printer according to correct connection scheme as described in Figure 3-7 on page 3-17.
  • Page 85: Connection Scheme: Color Printer

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-5-3-1 Connection Scheme: Color Printer Figure 3-7 Color Printer connection Chapter 3 - Setup Instructions 3-17...
  • Page 86: Connecting The Deskjet Color Printer

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-5-4 Connecting the DeskJet Color Printer 3-5-4-1 Connection via Bluetooth Adapter The DeskJet Color Printer can be connected to an external, non-isolated power source. The bluetooth adapter should be directly connected to the indicated USB 3 on rear panel Power Cable Bluetooth...
  • Page 87 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-5-4-1 Connection via Bluetooth Adapter (cont?) Please observe that the complete Bluetooth Printer Assembly has to be located outside of the CAUTION patient environment (according to IEC 60601-1 / UL 60601-1).
  • Page 88: Connecting The Wireless Network Adapter

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-5-5 Connecting the Wireless Network Adapter The Wireless Network Adapter can be connected to any accessible USB port of the Voluson™ P8 / Voluson™...
  • Page 89: Connecting The Footswitch

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-5-6 Connecting the Footswitch The footswitch should be directly connected to any accessible USB-port on the Voluson™ P8 / Voluson™ P6 (e.g., on rear of the system). NOTE: Connection of the Footswitch is always the same.
  • Page 90: Connecting The Usb Flash Memory Stick

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-5-7 Connecting the USB Flash Memory Stick NOTICE Before connecting an USB device, please read General Remarks and Hints when using external USB- Devices on page 3-23.
  • Page 91: General Remarks And Hints When Using External Usb-Devices

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-5-9 General Remarks and Hints when using external USB-Devices Do not connect or disconnect any external USB-devices to or from the system while WARNING WARNING scanning a patient! The appearing dialogs could distract you from the scan!
  • Page 92 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL To stop the external device, select it and then click the STOP DEVICE button. Figure 3-12 Device can now be safely removed Confirm the ‘Stop Device‘...
  • Page 93: Completing The Set Up

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 3-6 Completing the Set Up 3-6-1 Connecting the Unit to a Power Source The connection of the Voluson™ P8 / Voluson™ P6 ultrasound unit to a power source should be performed by a qualified person who has completed basic Voluson™...
  • Page 94 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Switch ON the Circuit at the rear of the system. Circuit Main Power Figure 3-13 Circuit and main power cable at rear of system 3-26 Section 3-6 - Completing the Set Up...
  • Page 95: Back End Processor Boot Up

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-6-2-2 Back End Processor Boot Up When AC power is applied to the Ultrasound unit, the ON/OFF button on the control panel illuminates NOTICE amber, indicating that the System (including the Back-end Processor) is in Standby mode.
  • Page 96: During A Normal Boot, You May Observe

    Figure 3-16 below. Figure 3-16 GE Healthcare wallpaper with progress bar 6.) The software initiates and sets up the Front-End electronics and the rest of the Ultrasound unit. 7.) The Keyboard backlight is lit. 8.) As soon as the software has been loaded, the 2D screen is displayed on the monitor.
  • Page 97: Power Off / Shutdown

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-6-3 Power Off / Shutdown NOTICE After turning off a system, wait at least 10 seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 98 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL NOTE: The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the ON/OFF button. So the auxiliary equipment need not to be switched ON/OFF separately. WARNING WARNING Disconnection of the Main Power Cable is necessary!
  • Page 99: Transducer Connection

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-6-4 Transducer Connection NOTE: When the probe is connected, it is automatically activated. Once connected, the probes can be selected for different applications. Connect a transducer to one of the three rightmost transducer receptacle as follows: 1.) Inspect the probe and probe socket to verify that it is free of debris.
  • Page 100: Printer Installation

    The Bluetooth Printer Connection set MUST NOT be installed by the user! Bluetooth installation please contact your local distributor or GE service representative. If a a Desk Jet printer(e.g., HP Officejet 100) is connected directly via an USB-cable, use the AC mains power outlet privided by the Voluson P-series system (auxiliary output).
  • Page 101: Installing The Network Color Laser Printer

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-7-3 Installing the Network Color laser Printer 1.) Power off/Shutdown the system as described in Section 3-6-3 on page 3-29 2.) Connect the printer as described in page 3-16 3.) Power On/Boot Up the system as described in...
  • Page 102 Keep in mind this unique printer IP Address (IPv4 Address). It has to be entered during Printer Installation Procedure. 11.)Download the appropreate printer driver from GE folders http://libraries.ge.com/foldersIndex.do? entity_id=52794606101&sid=101&SF=1 and extract files to CD or USB disk. 12.)Connect the USB stick to the Voluson P-series system or insert the CD into the drive.
  • Page 103 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 15.)On the right side of the screen select Connectivity and then click the Peripherals tab. 16.)Click the Add Printer button. Please read the displayed message carefully and click Yes if you have skills to do this.
  • Page 104 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL NOTE: This example show fictional numbers! Figure 3-26 enter printers IP Address 21.)Click on Have Disk and browse for E:\ (where E : stands for the USB stick resp. the CD drive), search for the printer driver folder and open the file hpcu175u.inf.
  • Page 105 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 22.)Select HP Universal Printing PCL6 and then Click OK to start the installtion. Figure 3-28 23.)Confirm the following screeen with Next. 24.)Remove the check mark at Set as the default printer and then click on Print a test page.
  • Page 106 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 25.)In the following window check mark No, do not send this info to HP and do not ask me again and then click OK.
  • Page 107: Installing The Deskjet Color Printer Directly Via An Usb-Cable

    2.) Connect the Bluetooth Adapter according to connection scheme, see 3.) Connect the USB "standard GE" service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up. Remove the dongle.
  • Page 108 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL NOTE: All software drivers are pre-installed for the designated printer only, see . If the Bluetooth device is already up and running, you can see it in the Device Manager, see . In this case, skip installation of the Bluetooth Adapter, close all open windows and proceed with Section 3-7- 5-2 on page...
  • Page 109 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10.)Select Browse my computer for driver software. Figure 3-36 Browse for driver 11.)Click on Browse and navigate to F:\(where F : stands for the DVD drive), select Win7 and click OK. Figure 3-37 Win7 12.)Verify the path and click Next to start installation.
  • Page 110 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 13.)When installation is finished click on Close. Figure 3-39 installation finished 14.)Now you should see the installed Bluetooth device in the Device Manager. Click the upper trackball key and verify its device status: “This device is working properly”...
  • Page 111: Prepare The Deskjet Color Printer

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-7-5-2 Prepare the DeskJet Color Printer NOTE: Preparations depend on the used Bluetooth Printer. HP Officejet 100 1.) Connect the power cord to a wall outlet socket. 2.) Turn on the printer.
  • Page 112 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8.) Select Add a network, wireless or Bluetooth printer. Figure 3-43 add Bluetooth printer 9.) Select your printer and click on Next. Figure 3-44 select printer 10.)You have successfully added the printer.Click Next.
  • Page 113 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 11.)Check box desired printer sharing option and then click Next. Figure 3-46 printer sharing 12.)Check box “Set as the default printer” and click Print a test page. Figure 3-47 print test page 13.)When test page is printed, click on Close and Finish.
  • Page 114: Hp Officejet 100 Turns Off Unexpectedly

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 14.)Verify correct printer settings;see 15.)Assign the printer as Report Printer, see 3-7-5-3-1 HP Officejet 100 turns off unexpectedly ( after being idle for a period of time) Issue A print, copy or scan job fails becuase the printer is turned off.
  • Page 115: Printer Installation Manually

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-7-6 Printer Installation manually 1.) Press the UTILITIES key on the control console. 2.) In the “Utilities” menu select SETUP button to invoke the setup desktop on the screen. 3.) On the right side of the screen select CONNECTIVITY and then click on the PERIPHERALS tab.
  • Page 116: Adjustment Of Printer Settings

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-7-7 Adjustment of Printer Settings 1.) After system restart, press the UTILITIES key on the control console. 2.) In the “Utilities” menu select SETUP button to invoke the setup desktop on the screen. 3.) On the right side of the screen select CONNECTIVITY and then click on the PERIPHERALS tab.
  • Page 117: Up-D897 / Up-D898Md - Printer Settings

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-7-7-1 UP-D897 / UP-D898MD - Printer Settings 1.) Call up the ‘Printer Preferences’; operation see: Section 3-7-7 "Adjustment of Printer Settings" 2.) Select the LAYOUT page and select:...
  • Page 118: Up-D25Md - Printer Settings

    Download the SonyUp-D25MD.reg from GE folders http://libraries.ge.com/foldersIndex.do?entity_id=15869258101&sid=101&SF=1 and extract files to CD or USB disk. • Connect the USB “standard GE” service dongle and enter the “Standard GE Revolving” password. • Open the Windows Start menu by pressing CTRL +ESC on the keyboard, click on “WinExplorer”...
  • Page 119 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) Color Correction. check mark Printer Hardware Color Correction Figure 3-53 Graphic page (lightness /Color Correction) 7.) Gray Balance. X = 50, Y = 50 8.) Adjustment by Color.
  • Page 120: Remote Control Selection

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-7-8 Remote Control Selection To assign an auxiliary device (e.g., printer) to the remote keys P1, P2, and/or P3 1.) Press the UTILITIES key on the control console. 2.) In the “Utilities”...
  • Page 121: Report Printer Selection

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-7-8-1 Report Printer Selection 1.) Click on the PERIPHERALS tab. (refer to:Figure 3-49: System Setup - Connectivity - PERIPHERALS page on page 3-47) 2.) Select the desired Report Printer from the pull-down menu.
  • Page 122: Cellular Modem Installation

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 3-8 Cellular Modem Installation 3-8-1 Overview This section describes the installation procedure of Cellular Modem Kit. 3-8-2 Tools Required Standard Field Service Tool Kit.
  • Page 123: Removal Procedure

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-8-4 Removal procedure 3-8-4-1 Removal procedure for Cellular Modem installation 1.) Remove the power cord and REAR CABINET BEXEL ASSY BT16. POWER CORD FIX Bracket REAR CABINET BEZEL ASSY BT16 Figure 3-57 Removing the POWER CORD and REAR CABINET BEZEL ASSY BT16...
  • Page 124: Installation Procedure

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-8-5 Installation procedure 1.) To refer to below picture, Unplug power cable for cellular modem and cut cable tie. Figure 3-59 Unpluging power cable 2.) Slide up the upper cover of the Cellular Modem(Pos.1)to get access to the SIM card holder.
  • Page 125 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Put in the SIM card like it is shown in the picture below, fix the SIM card with the mouted clips and put back on the upper cover.
  • Page 126 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) Refer to cellular modem guide and attach Cellular modem to REAR CABINET BEZEL ASSY BT16 like below. Figure 3-64 attaching Cellur modem to REAR CABINET BEZEL ASSY BT16 7.) Plug power cable and USB cable(to port 3) for cellular modem like below picture.
  • Page 127 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8.) Refer to below picture, insert two cables(USB and power for cellular modem) through hole of REAR EMI COVER. Figure 3-66 Inserting two cables through hole of REAR EMI COVER 9.) Assemble REAR EMI COVER and screw 7 screws.
  • Page 128 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10.)Pull out two cables from REAR EMI COVER First, pull out power cable within the limits of the possible. Therefore, pull out USB cable like below picture. To align cable edge, pull out USB cable as much as power cable.
  • Page 129 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 12.)Close REAR CABINET BEZEL ASSY BT16 and connect POWER COARD FIX BRACKET. POWER CORD FIX BRACKET REAR CABINET BEZEL ASSY BT16 Figure 3-70 Closing REAR CORD FIX BRACKET Chapter 3 - Setup Instructions 3-61...
  • Page 130: System Configuration

    • Annotation: adjustment of different Annotation presets(e.g. Home position, etc.) • Clipboard: adjustment of Clipboard display and functions • Service tab: conncet the USB “standard GE” dongle and enter the “standard GE revolving” password (change every six months_ to get access to the Service Tools functions.
  • Page 131: To Invoke The Setup Procedure

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-9-1-1 To invoke the Setup procedure: 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to activate the setup desktop screen. 3.) Select the corresponding major group from the left side of the screen and then click the desired tab.
  • Page 132: How To Enter Date And Time

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-9-1-2 How to enter Date and Time 1.) Invoke System Setup as described in Section 3-9-1-1 on page 3-63. 2.) On the left side of the screen select GENERAL SETTINGS and then click on the GENERALSETTINGS tab..
  • Page 133: How To Enter Clinic Name

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-9-1-3 How to enter Clinic Name 1.) Invoke System Setup as described in Section 3-9-1-1 on page 3-63. 2.) On the left side of the screen select GENERAL SETTINGS and then click on the GENERALSETTINGS tab.
  • Page 134: How To Activate System Login

    If screen is locked you have to enter the password to get full system control. If password is unknown click EMERGENCY. This enables standard - but limited - operation. NOTICE The System Login password cannot be reset by the user! Please contact your GE service representative. !! NOTICE: 5.) Click SAVE&EXIT to save Settings and exit System Setup.
  • Page 135: How To Change Ext. Monitor Output Settings

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-9-1-6 How to change Ext. Monitor Output Settings 1.) Invoke system Setup as described in Section 3-9-1-1 on page 3-63 2.) On the right side ofr the screen select Connectivity and then click the Peripherals tab.
  • Page 136: On-Board Optional Peripherals

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-9-2 On-Board Optional Peripherals Mains outlet: AC outlet for on-board peripherals Output voltage : 100~120/220-240VAC. Modification of voltage setting only by an authorized service person! CAUTION The maximum power consumption of equipment (inclusive of the color LCD monitor) connected !! CAUTION:...
  • Page 137: External I/O Connectors

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-9-3 External I/O Connectors Figure 3-74 External I/O Connectors - on Rear of System Table 3-3 External I/O Connector - Description Item Connector Name Table Number...
  • Page 138 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 3-5 RGB Out Connector, 19 Pin Pin No Output Signal Description RGB Output Greed RGB Output Blue RGB Output Horizontal - sync Horizontal -sync Vertical -sync...
  • Page 139: Available Probes

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 3-10 Available Probes Chapter 9 - Probes, on page 9-19, for part numbers to be used when ordering new or replacement service probes.
  • Page 140: Connectivity Setup

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 3-12 Connectivity Setup The Voluson™ P8 / Voluson™ P6 can be connected to various connectivity devices. The following sections describe how to connect the system to a remote archive/work station or a DICOM service, using a TCP/IP connection.
  • Page 141: Dataflow Examples

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-12-1-2 Dataflow Examples The local database is used for patient archiving. Images are stored to internal hard drive. The local database is used for patient archiving. Afterwards images are stored to a DVD/CD or external USB device, etc.
  • Page 142: Stand-Alone Voluson™ P8 / Voluson™ P6

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-12-1-3 Stand-alone Voluson™ P8 / Voluson™ P6 If digital images or 3D/4D data sets are stored, they should be saved in the Archive (Image Management System software).
  • Page 143: Network Configuration

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 3-13Network Configuration 3-13-1 TCP/IP Configuration NOTE: Following Information must be provided by customer or hospital engineer before you can start: A Station name, AE Title, IP address and Port Number for the Voluson™...
  • Page 144: Wireless Network Configuration

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL NOTE: This example shows fictional numbers! Figure 3-77 Internet Protocol (TCP/IP) You can get IP and DNS settings assigned automatically, if your network supports this capability. If fixed IP address is required, ask your network administrator for the appropriate settings.
  • Page 145: Disconnecting From The Wlan

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) The Wireless Network Configuration tool with available Wireless Networks appear. Figure 3-78 Wireless Networks - Connect 7.) Check box “Enable Wireless Connection”. 8.) Highlight the wireless network you want to use and then click Connect.
  • Page 146: Refreshing A Wlan Network

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Select the Security tab and then click Add. Figure 3-79 Security - Add 6.) Add the following information to the Wireless Network Properties page: Network Name(SSID) Check box “...
  • Page 147: Removing A Wlan Profile

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-13-2-6 Removing a WLAN Profile 1.) Press the “Utilities” key on the control console. 2.) On the right side of the screen select Connectivity and then click the Device Setup tab. 3.) Click the WLAN Configuration button;...
  • Page 148: Available Wlan Channels

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-13-2-8 Available WLAN Channels The available WLAN channels show availability of wireless connect point that the system can talk to. Each channel supports a finite number of users and has limited signal strength.
  • Page 149: Wlan Diagnostic

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Select the Monitor tab . Figure 3-82 Monitor 3-13-2-10 WLAN Diagnostic Running diagnostics is good if you think the adapterr might be bad. Diagnostics also tell you if the connection is behaving properly.
  • Page 150: Repairing The Wlan

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-13-2-11 Repairing the WLAN Occasionally you may need to repair a WLAN that has lost it’s connection to the Voluson™ P8 / Voluson™...
  • Page 151: How To Setup The Cellular Modem

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-13-3 How to Setup the Cellular Modem By means of the Cellular Modem, it is possible to connect the Voluson™ P8 / Voluson™ P6 to the Internet by using a SIM card.
  • Page 152 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Modem Configuration 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-29 2.) Insert the SIM card into the Cellular Modem. 3.) Connect the Cellular Modem as described in Section 3.4.5 on page 3-17 .
  • Page 153: How To Setup E-Mail

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-13-4 How to Setup E-mail If Internet access is available via LAN, Wi-Fi or Cellular modem, it is possible to send images/cines via E-Mail to patients or physicians.
  • Page 154: How To Setup E-Mail To Mms Service

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-13-5 How to Setup E-mail to MMS Service If the Voluson™ P8 / Voluson™ P6 is connected to the Internet (via LAN, Wi-Fi or Cellular modem) and E-mail service is configured properly, it is possible to send images as an MMS.
  • Page 155: How To Enter Patient's Email Address And Phone Number In The Pid Screen

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8.) Individually enter Patient's Email address and Phone number in the PID screen and adjust Email and/or MMS options as described in Figure 3-13-6 on page 3-87 3-13-6 How to enter Patient's Email address and Phone number in the PID screen...
  • Page 156: Map Network Drive

    RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3-13-7 Map Network Drive 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select CONNECTIVITY and then click the DRIVES tab.
  • Page 157: Insite Exc Configuration

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) Select the CONNECT button to establish the connection to the remote machine. If successful, the DISCONNECT button becomes active. NOTICE If there is an error during the connection, a warning message appears inside the dialog.
  • Page 158 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Select the Configuration page, then double-click InSite Exc Agent Configuration(A). Figure 3-91 Common Service Desktop - Configuration NOTE: This example shows fictional numbers! 3.) Fill out at least all bold stated mandatory fields in the Agent Configuration section(B).
  • Page 159: How To Create An Insite Permanent User

    An InSite permanent user is required for automatic system error reporting to the digital service network. Moving the cursor to the InSite ExC link(GE icon) at the right bottom of the display screen and press the left trackball key(=left-click). The “Contact GE”...
  • Page 160 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4. In the next screen click Add user(3). Figure 3-93 Add User 5. Fill in the required information and confirm with Add User(4). Figure 3-94 Confirm with Add User 6.
  • Page 161: Connectivity Setup Worksheet

    DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 3-14 Connectivity Setup Worksheet Site System Information Comments: Floor: Site: Dept: Room: Voluson™ P8 / Type: Voluson™ P6 REV: CONTACT INFORMATION Name Title...
  • Page 162 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 3-14 Connectivity Setup Worksheet (cont?) Voluson™ P8 / Voluson™ P6 IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination...
  • Page 163: Paperwork

    Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. Chapter 3 - Setup Instructions...
  • Page 164 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 3-96 Section 3-15 - Paperwork...
  • Page 165 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the Voluson™ P8 / Voluson™ P6 System diagnostics by using the built-in service software, and power supply adjustments.
  • Page 166 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. !! CAUTION: The covers are required for safe operation, good system performance and cooling purposes.
  • Page 167 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-3-1-1 System Power On (cont?) 5.) Hold down the On/Off button (see: Figure 4-2) on the control panel for ~3 seconds. NOTE: The power for on-board peripheral auxiliary equipments are commonly switched with the ON/OFF button.
  • Page 168 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-3-2 Power Off / Shutdown NOTICE After turning off a system, wait at least 10 seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 169 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) After complete power down, unscrew the 2 screws and remove the pull-out protection to disconnect the main power cable from the system or unplug it from the AC wall outlet socket. Circuit breaker Pull-out Protection 2 screws...
  • Page 170 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-3-3 System Features 4-3-3-1 Control Panel Figure 4-5 Control Panel Tour 1. Loudspeaker position 7. Menu window flipswitch controls 2. Menu window digipot controls 8.
  • Page 171 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-3-3-2 Monitor Display Figure 4-6 Monitor Display Tour Table 4-2 Monitor Display Features 1. Logo 15. Measurement results 2. Patient Name (Last-, First-, Middle Name) 16.
  • Page 172 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 4-4 Functional Checks For a basic functional check of the system’s different modes, refer to the Voluson™ P8 / Voluson™ P6 Basic User Manual.
  • Page 173 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-4-1 Patient Archive (Image Management) The Voluson™ P8 / Voluson™ P6 provides an Patient/Image Management System that allows fast and extremely easy patient, exam and image management.
  • Page 174 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-4-2 Erasing DVD/CD 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select CONNECTIVITY and then click the DRIVES tab.
  • Page 175 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 4-5 Backup and Restore Database, Preset Configurations and Images It is recommended to Backup the Full System Configuration (Section 4-5-3 on page 4-16) and the CAUTION Image Archive...
  • Page 176 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-5-1 Save Small Backup (Scan Settings) The Image Settings contains: • Application Settings • 2D Factory and 2D User Presets •...
  • Page 177 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-5-2 Load Small Backup (Scan Settings) The loading procedure overwrites the existing image settings on the local hard drive. CAUTION Make sure to insert the correct System DVD.
  • Page 178 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-5-2-2 Load “Complete Backup” NOTE: Following procedure should be used, if the user prefers factory default settings, which are adapted for the installed Application Software version.
  • Page 179 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-5-2-3 Load only parts of the “Complete Backup” NOTE: Following procedure should be used, if the user has saved own 2D/3D/4D Programs or AutoText entries, but new presets have to be added to match the installed Application Software version (e.g., 2D/ 3D/4D Programs for new probe, etc.).
  • Page 180 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-5-3 Save Full System Configuration (Full Backup) A backup of the Full System Configuration always contains the following data • User Settings (databases and files containing User Programs, 2D/3D/4D Programs, Auto Text entries, gray curves and complete System settings such as language, time/date format, etc.) •...
  • Page 181 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-5-3 Save Full System Configuration (Full Backup) (cont?) When the “Full Backup” is saved on a network drive (to map a network drive see: Section 3-13-2 "Wireless Network Configuration"...
  • Page 182 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-5-4 Load Full System Configuration (Full Backup) It is recommended to backup data before an upgrade; see: Section 4-5-3 on page 4-16.
  • Page 183 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-5-4 Load Full System Configuration (Full Backup) (cont?) 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select BACKUP and then click the SYSTEM CONFIGURATION tab.
  • Page 184 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-5-4 Load Full System Configuration (Full Backup) (cont?) 9.) Click the NEXT button and then select YES to start, or NO to cancel the restore procedure. Figure 4-21 Start Restore Backup now? After restoring the data, the Voluson™...
  • Page 185 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-5-6 Archiving Images It is recommended to Save Image Archive 4-21) and Backup the Full (Section 4-5-6-1 on page CAUTION System Configuration 4-16) once a week.
  • Page 186 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-5-6-1 Save Image Archive (cont?) 5.) Choose archiving span time from the pop-up menu. If for example “All images older than” 1 Day is chosen (see NOTICE Figure 4-24 on page 4-21), images of the...
  • Page 187 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-5-6-2 Load Image Archive 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select BACKUP and then click the IMAGE ARCHIVE tab.
  • Page 188 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 4-6 Software Configuration Checks 4-6-1 System Setup Press the F2 key on the control panel and then select SYSTEM SETUP. On the left side of the screen select the desired major group.
  • Page 189 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-6-2 Measure Setup Press the F2 key on the control panel and then select MEASURE SETUP. The Measure Setup desktop offers three different pages to check. NOTICE Parameters and possible adjustments depend on the selected Application.
  • Page 190 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 4-7 Peripheral Checks Check that peripherals work as described below: Table 4-7 Peripheral Checks Step Task to do Expected Result(s) Press the FREEZE key.
  • Page 191 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-7-2 Control Console Positioning The control console can be adjusted to a fixed height (910 mm) with an Extender Kit (H46712CE). 4-7-2-1 Assembly procedure for converting Voluson™...
  • Page 192 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Unscrew 2 screws to remove the MONITOR CABLE COVER. And unscrew 5 screws to disconnect the LCD cables. Figure 4-30 Removing the MONITOR CABLE COVER and cables 4.) Unscrew 1 screw to remove MONITOR ARM COVER BT16.
  • Page 193 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) Unscrew 9 screws under the keyboard base. Figure 4-33 Unscrewing 9 screws under the keyboard base 7.) Remove 5 Probe holders from the keyboard assembly. Figure 4-34 Removing 5 probe holders Chapter 4 - Functional Checks 4-29...
  • Page 194 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8.) Remove keyboard assembly. Pull it downwards Figure 4-35 Removing lower keyboard assembly 9.) Disconnect the OPIO cable, Power cable, USB cables from keyboard assembly. Figure 4-36 Disconnecting the cables from keyboard assembly 4-30 Section 4-7 - Peripheral Checks...
  • Page 195 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10.)Unscrew 4 screws to remove upper keyboard assembly. Figure 4-37 Unscrewing 4 screws to remove upper keyboard assembly 11.) Disconnect the OPIO cable, power cable, USB cables from upper keyboard assembly Figure 4-38 Disconnecting the OPIO cable, Power cable and USB cable from upper keyboard assembly Chapter 4 - Functional Checks 4-31...
  • Page 196 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 12.)Remove the cable tie. Cable tie 2160955 Figure 4-39 Removing cable tie 13.)Pull the LCD cables through the LCD ARM and MONITOR HOLDER. Figure 4-40 Pulling LCD cables through the LCD ARM 14.)Disconnect Speaker cables.
  • Page 197 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 15.)Unscrew 2 screws to remove the ground cables . ground cables, remove 2 screws Figure 4-42 Removing the ground cables 16.)Unscrew 4 screws to disassemble the keyboard base from the pedestal.
  • Page 198 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 18.)Assemble Extender with pedestal using 4 flat washers, spring washers and hexagon socket head cap. Refer to figure below. Figure 4-45 Assembling Extender Table 4-8 Part list of Extender GEPN...
  • Page 199 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 20.)Assemble the keyboard base using 4 flat washers, 4 spring washers and 4 hexagon socket screws. Figure 4-47 Assembling the keyboard base 21.)Re-assemble LCD cables through the LCD ARM, Refer to figure below.
  • Page 200 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 23.)Screw 1 screw to assemble MONITOR ARM COVER BT16. Then screw 5 screws to connect the LCD cables. Figure 4-50 Screwing 1 screw to assemble LCD ARM COVER then connect the LCD cables 24.)Connect the speaker cables speaker cables Figure 4-51 Connecting the speaker cables...
  • Page 201 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 25.)Connect the ground cable ground cables, Screw 2 screws Figure 4-52 Connnecting the ground cables 26.)Connect the power cable and USB cables power cable USB cables Figure 4-53 Connecting the power and USB cables...
  • Page 202 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 29.)Screw 2 screws and assemble the upper keyboard assembly. Figure 4-55 Screwing 2 screws and assembling the upper keyboard assembly 30.)Connect OPIO, power and USB cables.
  • Page 203 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 32.)Screw 9 screws under the KEYBOARD BASE. Figure 4-58 Screwing 9 screws under the KEYBOARD BASE 33.)Reassemble 5 probe holders Figure 4-59 Assembling 5 probe holders 34.)Screw 2 screws to assemble SPEAKER COVER LEFT AND RIGHT.
  • Page 204 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 35.)Assemble 2 RUBBER CAP on SPEAKER COVER LEFT/RIGHT. Figure 4-61 Assembling 2 Rubber cap on speaker cover left/right 36.)Screw 2 screws to assemble MONITOR CABLE COVER. Figure 4-62 Screwing 2 screws to assemble MONITOR CABLE COVER 4-40 Section 4-7 - Peripheral Checks...
  • Page 205 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-7-3 Brakes and Direction (Swivel) Locks Check the brakes and swivel locks function as described below. Table 4-9 Brakes and Direction (Swivel) Lock Step Task Expected Result(s)
  • Page 206 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 4-8 Mode Transition 4-8-1 B Mode 1.) Make sure that the probe is connected properly. 2.) Select the probe on the screen menu. 3.) Press [B] key on the operator panel to enter only B mode.
  • Page 207 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 4-9 Site Log Table 4-10 Voluson™ P8 / Voluson™ P6 - Site Log (Paper Documentation)Voluson™ P8 / Voluson™ P6 Date Service person Problem...
  • Page 208 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4-44 Section 4-9 - Site Log...
  • Page 209 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains Voluson™ P8 / Voluson™ P6’s system concepts, component arrangement, and subsystem function.
  • Page 210 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 5-2 General Information Voluson™ P8 / Voluson™ P6 is a digital beamforming curved-, linear- and phased array ultrasound imaging system. The system can be used for: •...
  • Page 211 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 5-2 General Information (cont?) Among other significant features of the Voluson™ P8 / Voluson™ P6 are the following: Integrated FrontEnd (uses advanced FGPA technologies) Open connectivity using USB ports Bluetooth - wireless connectivity High performance 17”...
  • Page 212 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 5-2 General Information (cont?) Figure 5-2 Basic Block diagram of Voluson™ P8 / Voluson™ P6 Section 5-2 - General Information...
  • Page 213 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 5-2 General Information (cont?) The Voluson™ P8 / Voluson™ P6 used digital beamforming technology which provides high resolution and high penetration performance. It is a general purpose, mobile, software controlled diagnostic ultrasound System .
  • Page 214 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-1 Description of Voluson™ P8 / Voluson™ P6 Operating Modes 5-2-1-1 B-Mode or 2D-Mode B-Mode or 2D-mode is a two-dimensional image of the amplitude of the echo signal. It is used for location and measurement of anatomical structures and for spatial orientation during operation of other modes.
  • Page 215 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-1-3 Color Doppler Modes Color Doppler is used to detect motion presented as a two-dimensional display. There are following applications of this technique: •...
  • Page 216 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-1-4 Pulsed (PW) Doppler PW Doppler processing is one of two spectral Doppler modalities. In spectral Doppler, blood flow is presented as a scrolling display, with flow velocity on the Y-axis and time on the X-axis.
  • Page 217 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-1-8 3D Rendering The 3D (volume) rendering is a calculation process to visualize certain 3D-structures of a scanned volume by means of a 2D-image. The gray value for each pixel of the 2D-image is calculated from the voxels along the corresponding projection path (analyzing beam) through the volume.
  • Page 218 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-2 Block diagram Voluson™ P8 / Voluson™ P6 Figure 5-4 Voluson™ P8 / Voluson™ P6 - Block diagram 5-10 Section 5-2 - General Information...
  • Page 219 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-2-1 Front End Block diagram For further details : see: "Front End Processor - Block diagram" on page 5-22 see: "T-DPI - Blockdiagram" on page 5-23 see: "Beamformer + Frontend"...
  • Page 220 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-3 Data Flow Control Description This section describes the functions of the boards vs. different operation modes. • DBM64 - Beamformer Motherboard •...
  • Page 221 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-3-1-1 Special B-Mode Techniques a.) HI (Coded Harmonic Imaging): In one method of HI the RX-Frequency is doubled, so that the radial resolution is increased due to the higher RX-Frequency.
  • Page 222 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-3-4 D-Mode Autotrace (draws PC-calculated envelope to D-Spectrum) 1.) DRFM PRF-generator; see: Section 5-2-3-1 "B-Mode" on page 5-12 After DC-cancelling the signal is mixed with RX-Frequency and brought to LF-Spectrum, where the LowPassFilter cuts HF.
  • Page 223 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-3-5 CFM-Mode (Color Flow Mode) 1.) DRFM PRF-generator; see: Section 5-2-3-1 "B-Mode" on page 5-12 After DC-cancelling the signal is mixed with RX-Frequency and brought to LF-Spectrum, where the LowPassFilter cuts HF.
  • Page 224 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-4 Description of Software Options To activate the software options: 1.) Press the F2 button on the OPIO. 2.) On the left side of the screen select SYSTEM SETUP to activate the setup desktop screen. 3.) On the left side of the screen select ADMINISTRATION and then click the OPTIONS tab.
  • Page 225 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-4-1 3D/4D Activation 5-2-4-1-1 4D Real Time 4D Real Time mode is obtained through continuous volume acquisition and parallel calculation of 3D rendered images.
  • Page 226 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-4-5 SingleView More flecxibility with any Plane, VCI plane is freely selectable. Any shape can be drawn. • Volumes can be editaed in all other visualization Modes. •...
  • Page 227 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-4-12 SonoNT SonoNT is an additional function for manual NT (Nuchal Translucency) measurement. This function supports the user to find the correct position for the NT measurement. The user can switch between NT Method “Manual”...
  • Page 228 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-5 Description of Hardware Options Table 5-3 Hardware Options HW-Options Description 5-2-5-1 CW - Continuous Wave Doppler CW-Doppler 5-2-5-2 ECG Module ECG Module 5-2-5-3 Wireless Network Adapter (WLAN - Wireless Local Area Network) WLAN Network Adapter...
  • Page 229 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-2-6 Data Location The Voluson™ P8 / Voluson™ P6 has a 500 Gbyte Hard disk divided into 4 partitions: C: System partition: Operating System (Windows 7) including all Windows settings (IP-address, Network Name, etc.) US-Application Software (UISAPP)
  • Page 230 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 5-3 FrontEnd Processor Figure 5-6 Front End Processor - Block diagram Table 5-4 below outlines the Voluson™ P8 / Voluson™ P6 Front End components described in the sub- sections.
  • Page 231 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-3-1 T-DPI - Probe Interface board Switches the Probe Connectors (3 RS -Connectors) and recognizes Probes • 3 RS-Probe Connectors •...
  • Page 232 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-3-2 DBM64- Beam former board DBM64DBM64 board has 64 TX/RX channels, respectively. beamformer board is a complete 64 channel ultrasound Front-end that includes the following features and functions:. TX-pulse genetration, TX-Apodization, TX-Focus RX-Focus Delay and summation Full receiving channel with integrated AFE IC: TGC + Anti-Aliasing Filter + ADC...
  • Page 233 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-3-3 DRFM- Radio Frequency interface and System controller Figure 5-9 DRFM - Block diagram Chapter 5 - Components and Functions (Theory) 5-25...
  • Page 234 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-3-3-1 DRFM - FPGA A.) Beamformer interface • LVDS-interface • Raw ultrasound data is transferred from the Beamformer to the Interface FPGA. •...
  • Page 235 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 5-4 BackEnd Processor Figure 5-10 BackEnd Processor - Block diagram Chapter 5 - Components and Functions (Theory) 5-27...
  • Page 236 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 5-5 Voluson™ P8 / Voluson™ P6 - DRFM components Sub-section Description Page Number 5-4-1 5-28 Hard Disk Drive 5-4-2 5-28 SOM (System On Module) 5-4-1...
  • Page 237 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Control DVD drive (SATA), User Interface (USB), DVR (USB) Figure 5-11 SOM (System On Module) Chapter 5 - Components and Functions (Theory) 5-29...
  • Page 238 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 5-5 Internal I/O Figure 5-12 Voluson™ P8 / Voluson™ P6: Internal I/O 5-30 Section 5-5 - Internal I/O...
  • Page 239 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 5-6 OPIO (User Interface) 5-6-1 OPIO Block Diagram A/N Key USB Cable 8 wire (USB (4) + LED backlight) USB HUB Down Stream(2) * Stanby 3.3V...
  • Page 240 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-6-2 OPIO layout 1. Loudspeaker position 2. Menu window digipot controls 3. TGC Slider controls 4. Trackball 5. Alphanumeric Keyboard 6.
  • Page 241 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Chapter 5 - Components and Functions (Theory) 5-33...
  • Page 242 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Programmable key 5-34 Section 5-6 - OPIO (User Interface)
  • Page 243 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Function of the Trackball at Diverse Dialog Pages In general operations at diverse dialog pages and windows on the system desktop (e.g., Patient Data Entry, Usage of EUM, System Setup, Measurement Setup, etc.) are executed by trackball and the trackball keys (mouse emulation).
  • Page 244 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL SW-DVR Go to SW-DVR menu 5-36 Section 5-6 - OPIO (User Interface)
  • Page 245 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 5-7 Monitor NOTE: From Voluson™ P8 / Voluson™ P6 system, it's accessible via not monitor adjustment but monitor menu on application.
  • Page 246 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 5-9 External I/O Figure 5-16 External I/O Connectors - on Rear of System (GES) Item Connector Name Description USB-2.0 port Connector for external Monitor HDMI OUT TMDS signal with DDC_Control (HDMI or DVI)
  • Page 247 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 5-10 Peripherals 5-10-1 Printers 5-10-1-1 Black & White Digital Printer The B&W Digital Printer receives image data via the USB port. The print command is controlled by the keys P1, P2, or P3 on the Voluson™...
  • Page 248 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 5-11 Power Distribution 5-11-1 ACFE - Primary Power Module Curcuit breaker Equipotential Connection Main Voltage Indicator (Shows input voltage level) Pull-out Protection Main power cable Figure 5-18 Primary Power Module...
  • Page 249 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-11-1-1 DPS Block Diagram Figure 5-19 DPS Block Diagram Chapter 5 - Components and Functions (Theory) 5-41...
  • Page 250 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-11-1-2 Power Distribution Block Diagram T-DPI Figure 5-20 Power Distribution Block Diagram 5-42 Section 5-11 - Power Distribution...
  • Page 251 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-11-1-3 Power Up Sequence Description The Power Up Sequence can be divided in the following steps: • Switch Circuit Breaker in DPS to ON position •...
  • Page 252 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-11-1-3-2 Power Down Sequence Description Figure 5-22 Power Down Sequence 1.) On/Off Switch is pressed on OPIO. 2.) ON/OFF signal is sent from On/Off Switch via OPIO cable to DRFM 3.) “Shut down”...
  • Page 253 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-11-2 ACDC - primary DC Power Supply 5-11-2-1 Function of ACDC With the AC input voltage from the ACFE, the ACDC module generates 12Vdc 5-11-3 DCDC - main DC Power Supply With the 12Vdc from ACDC, the DCDC module generates all DC power sorces:...
  • Page 254 GE to deliver remote diagnostics capability. InSite is your direct link with a GE Online Service Engineer or Applications Support Engineer, or a Request for Service via the Insite Exc link ( GE icon) at the right botton of the display screen.
  • Page 255 3.) In the Utilities menu select SETUP button to invoke setup desktop on the screen. 4.) Connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up. Remove the dongle.
  • Page 256 NOTE: ONLY registered GE Service personal have access to this feature. InSite ExC allowas GE technicians to view the entire customer’s desktop and operation system. Using VCO(Virtual Console Observation) a service technician or the OLC(OnLine Center) can access and modify all settings and programs or run diagnostics on the customer’s Voluson P-Series system.
  • Page 257 5-13-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. There are different possibilities to access the Common Service Desktop and its available features:...
  • Page 258 2.) In the “Utilities” menu select SETUP button to invoke the setup desktop on the screen. 3.) Connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up.Remove the dongle.
  • Page 259 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-14-3-1 Auto Tester Autotester is a log function of customer activities. It records all user actions (scanning, entries, performing Calculations, review of Patient Reports, etc.…). It is possible to safe (record) as file on HDD. But also export to DVD/CD can be done to allow replay of the records on other units.
  • Page 260 Each user is only granted access to the tools that are authorized for thier use. 5-14-3-6 Request for Service(DRFM) Fill out the “Request For Service” form and then send the problem description to GE Service/Application representatives. Operation see: Section 7-2-1 "Request for Service (DRFM)" on page 7-4...
  • Page 261 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-14-3-8 Export System Data Select the EXPORT SYSTEM DATA button on the “Service Tools” page to Full Backup the System State.
  • Page 262 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-14-3-13 Load Application Settings If the Tune version of the Application presets does not match the Application Software version, it is probably that there are adverse affects on image quality (e.g., after reloading an old “Full Backup”).
  • Page 263 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-14-3-14 Probe EEPROM NOTE: It is recommended to connect and re-program one probe after the other. Do not assign wrong coding (e.g., programming a RAB to a RIC probe). An inaccurate probe CAUTION programming will mostly likely damage the probe and/or system.
  • Page 264 2.) As soon as the “Boot Screen” appears (see: Figure 5-31 down] key on the keyboard until the GE Service item is highlighted, then press ENTER. (case sensitive) and confirm with ENTER. The system boots into 3.) Enter the Password Rudi LINUX.
  • Page 265 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-15-1-3 Rollback This function offers the possibility to simply restore the previously used system configuration (rollback), which was stored on R:\ during “FMI from DVD”. 1.) Turn system OFF and then back ON.
  • Page 266 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-15-1-4 Memtest(Memory Check) 1.) Turn system OFF and then back ON. 5-56), press the PG DN 2.) As soon as the “Boot Screen” appears (see: Figure 5-31 on page [...
  • Page 267 ANUAL Section 5-16Windows Start Menu GE field engineers can find frequently used tools within a sub folder of the Windows Start menu. 1.) Turn power on (let system start up completely). 2.) Connect the USB “Standard GE” service dongle to the USB connector(s) on the rear of the system.
  • Page 268 DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 5-6 Sub folder “GE Field Engineer” Windows Tool Description Remap Drive Letters Automatically reassigns drive letters: C: System, D: User, F:CD-Drive, R:Rescue Service Manual Displays the Proprietary Service Manual of the system show hidden devices Device Manuager shows all devices ( e.g.
  • Page 269 • * via Boot Screen Functions - GE Service: functions are secured by service password (class C) * via Windows Start Menu (for GE Field Engineers ONLY): by means of service dongle and GE password System is NOT responsive: (booting from System DVD is required) •...
  • Page 270 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-17-3 Before Recovering or Initializing the Hard disk If the system is still responsive (system starts up without problems and the application software CAUTION is running as usual, try to print out or write down System Setup and Measure Setup settings;...
  • Page 271 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) On the right side of the screen select CONNECTIVITY: - Click the Device Setup tab and Click the NETWORK CONFIGURATION button and print out the “internet Prodocol (TCP/IP) Properties”...
  • Page 272 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-17-4 Voluson Maintenance System - Main Menu Enter Linux as described in Section 5-17-2 "Access/Security" on page 5-61. The “Voluson Maintenance System” main menu with following functions appear: Figure 5-35 Voluson Maintenance System - Main menu 5-17-4-1 Media...
  • Page 273 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-17-5 Media Menu The “Media Menu” is only accessible if the System DVD (see: Table 9-13) is inserted into the system’s NOTICE DVD/CD+R/RW Drive 1.) If system functionality allows it, collect vital system information as described in...
  • Page 274 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 1.) Select [Restore] in the displayed “Media Menu” (see: Figure 5-36 above). NOTE: If the message: “Recovery system (Linux) needs to be updated ...” appears, confirm with YES. NOTE: If the currently installed software has to be upgraded to a newer version, a new software specific “Permanent key”...
  • Page 275 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Please wait until the final window (see figure below) appears and then press Enter. Figure 5-40 Restore C; done 4.) In the displayed menus select [Exit] -> [Exit/Reboot]. Figure 5-41 Exit ->...
  • Page 276 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8.) If prompted, enter the proper “ Permanent key” and then click Submit. Figure 5-43 enter Permanent key NOTE: Options will be recovered at a later step using “Full Backup” process. 9.) Customize the system as described in Section 5-17-8 on page 5-80.
  • Page 277 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-17-5-3 NewDisk NOTE: The “NewDisk” function is ONLY available when booting from System DVD. This function performs operations which result in complete data loss of all files on the entire hard disk.
  • Page 278 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-17-5-3-1 BIOS “Boot Priority Order” Setting General operation and selection in BIOS and LINUX menu is done with: • the alphanumeric keyboard •...
  • Page 279 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) Press the [-] key on the keyboard to change the “Boot Priority” order: - 1st Boot: from Hard Disk (SATA 2 Drive) to DVD-Drive (SATA 0 Drive) - 2nd Boot: from DVD-Drive (SATA 0 Drive) to Hard Disk(SATA 2 Drive).
  • Page 280 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-17-5-3-2 New Hard Disk Setup NOTE: If not already done, change BIOS “Boot Priority Order” as described in Section 5-17-5-3-1 on page 5-70.
  • Page 281 9.) Using the “Arrow” keys [<-] or [->] on the alphanumeric keyboard select the [Save&Exit]tab. 10.)Select Load GE BIOS Setting xxxx (“xxxx” is the installed BIOS version) and then press Enter. This changes the “Boot Priority” order back to 1.Boot:Hard Disk.
  • Page 282 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 13.)Please wait until all processes are finished and the final “DOC” window (see figure below) appear. Figure 5-49 press the “A” key to start UlSApp 14.)Press A key to activate the Ultrasound Application Software [UIS].
  • Page 283 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-17-6 Advanced Functions Menu 1.) If system functionality allows it, collect vital system information as described in Section 5-17-8 "After Recovering or Initializing the Hard disk"...
  • Page 284 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-17-6-5 Initialize Hard (C: and D:) This function performs operations which result in complete data loss of all files on C:\ and D:\ partition.
  • Page 285 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-17-7 ServiceTools 1.) Enter Linux as described in 2.) In the “Voluson Maintenance System” main menu (Figure 5-35 on page 5-64) select [Advanced].
  • Page 286 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-17-7-3 USBTree The USB Tree displays connected devices and their respective hub. This enables easy control, whetherall devices are correctly wired. Figure 5-53 USB Tree (example) 5-17-7-4 SmartHDDtest...
  • Page 287 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5-17-7-6 CollectLogs Voluson Log files can be stored on an USB-stick. The stored zip file includes the current system state (as “Full Backup”), Process Logger files, Auto tester files, SMART logs, dump files, etc.
  • Page 288 1.) If not currently done, restart the system (with new HDD) and boot up completely. 2.) Connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up.
  • Page 289 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 This chapter describes how to test and adjust the mechanical capabilities of a System that may be out of specification.
  • Page 290 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 6-3 LCD Monitor Adjustment NOTE: From Voluson™ P8 / Voluson™ P6 system, it's accessible via not monitor adjustment but monitor menu on application.
  • Page 291 2.) In the “Utilities” menu select SETUP button to invoke the setup desktop on the screen. 3.) Connect the USB “Standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up. Remove the dongle.
  • Page 292 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL NOTE: If the desired language is not listed, click the ADD button, choose the desired input language from the drop down menu, as shown in Figure 6-2 below, and then confirm with OK.
  • Page 293 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter Voluson™ P8 / Voluson™ P6 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
  • Page 294 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 7-2 Collect Vital System Information 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select ADMINISTRATION and then click the SYSTEM INFO tab.
  • Page 295 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 7-2 Collect Vital System Information (cont?) Move the scroll bars downwards to view additional information about installed software/hardware (e.g., Console information). Figure 7-1 System Setup - Administration - SYSTEM INFO page (e.g., VP8) Chapter 7 - Diagnostics/Troubleshooting...
  • Page 296 !! NOTICE: There are two possibilities to contact GE: • by means of the GE “Remote Status Icon” that is displayed on the bottom of the screen. Move the cursor to the GE icon and press the left trackball key (= left-click) to display the “Request For Service”...
  • Page 297 5.) Write down and keep the Reference Number for follow up procedures, then click OK. If the service platform is not configured an Error message is displayed. The request is NOT sent! NOTE: The request is saved in QUEUE for reviewing and/or resending. Figure 7-4 Contact GE - Queue Chapter 7 - Diagnostics/Troubleshooting...
  • Page 298 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-2-2 Shortcuts List Press the CTRL + H key simultaneous to display the shortcuts list and a description of what they do. Figure 7-5 Shortcut list (Ctrl + H) Shortcut Description...
  • Page 299 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 7-3 Check Points Voltages 7-3-1 How to check power 7-3-1-1 Dachstein Power Supply (DPS) Turn on mains switch and check green LED of P5V, N5V, LIVE5V and 12V on backplane. Figure 7-6 Check green LED on backplane NOTE: If LED is on, DPS has full function (all voltages are within the specified range).
  • Page 300 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 7-4 Screen Captures and Logs There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center.
  • Page 301 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) On the network sniffer screen select Capture -> Options. Figure 7-8 Capture -> Options 3.) Choose your Network in the “Interface” pull-down menu (1) -> Intel Network or Wireless adapter. NOTE: Remember, if the device has a DVR, it will show up in the “Interface”...
  • Page 302 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL images to the storage device or query Worklist.) 8.) Press Alt + N to restore Sniffer window and observe the network activity. Packnet List”...
  • Page 303 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 11.)Select File -> Save As, enter the file name D:\export\xxxx (xxxx= name of your preference). Figure 7-10 Save captured values If you have used a screen filter , select “Displayed”(B).
  • Page 304 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-4-2-1 How to use a Capture Filter Use filters to limit the captured data. Filter by IP and port number by IP only.
  • Page 305 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-4-2-2 How to use a Screen Filter In addition to the “pre” capture filter, use the Filter tool in the screen to filter what is displayed. Filter the DICOM packets, since they are the most probable for the troubleshooting.
  • Page 306 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-4-3 Export Log’s and System Data There are two possibilities to export system data (and log´s): 1.) by pressing the ALT + D key to save a snapshot and “Full Backup” of the System state; see: Section 7-4-3-1 on page 7-14 SERVICE...
  • Page 307 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Figure 7-15 System Configuration Save 3.) Select the destination of the “System state” backup. 4.) Select the NEXT button to start the backup process. After saving the data, the Voluson™...
  • Page 308 2.) In the “Utilities” menu select SETUP to invoke the setup desktop on the screen. 3.) If not already done, connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up.
  • Page 309 How the Customer enables/disables Disruptive Mode and VCO 1.) If not already in read mode, FREEZE the image. 2.) Move the cursor to the InSite ExC link (GE icon aht the right bottom of the display screen) and press right trackball key (= right-click).
  • Page 310 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-5-3 Virtual Network Computing (VNC) NOTICE The system may require a correct proxy configuration. Please ask the customer. !! NOTICE: 1.) In the Total Access Session window (see: Figure 7-26 on page 7-24) find the item labeled...
  • Page 311 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) The next window shows the command to be issued, click on [OK] to grant the request. The “virtual” Monitor of the Voluson™ P8 / Voluson™ P6 is displayed on your remote workstaticcreen.
  • Page 312 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Press the CTRL + H key on the alphanumeric keyboard, to display a list of shortcuts available whenusing the “Virtual Network Computing (VNC)” feature. Figure 7-21 Tool bar on top of virtual screen TIP: Frequently used shortcut functions:...
  • Page 313 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Press the shortcut ALT + S on the alphanumeric keyboard, to display the Console Emulator. Figure 7-22 Voluson™ P8 / Voluson™ P6 Console Emulator NOTE: For description of keys and their functions refer to Section 5-6-2 "OPIO layout"...
  • Page 314 3.) The next window shows the command to be issued, click on OK to grant the request. Figure 7-24 allow command 4.) Select “GE Service” from the pull-down menu, enter the password (changes every six months) and then click Okay.
  • Page 315 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) If not already done, connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up.
  • Page 316 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 11.)Start scanning. You can scan normally and everything will be recorded to the program (up to several hours.) NOTE: It is important that you are recording the processes where the errors normally occur. Stop the program by clicking on [Stop] shown on the screen, or by pressing the ALT GR key on the alphanumeric keyboard.
  • Page 317 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-6-1 Limitation of the Auto Tester The following information will not be recorded. Depending on the moment the Auto Tester is activated, this must be provided by the customer or the field engineer.
  • Page 318 2.) In the F2 menu, press SYSTEM SETUP button to invoke the setup desktop on the sceen. 3.) If not already done, connect the USB “standard GE” service dongle to the USB connector(s) on the system. As soon as the dongle is detected, the login window pops up.
  • Page 319 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Automatic/Not-Automatic diagnostic tests are executed from this System Diagnostics Tool. Figure 7-30 System Diagnostics Tool Chapter 7 - Diagnostics/Troubleshooting 7-27...
  • Page 320 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-7-1 Audio(System) Interface Test. Test of the Audio path from system to Speaker and AudioOutput can be performed. 1.) Select the AUDIO(SYSTEM) INTERFACE TEST button in the Diagnostics Tool window. 2.) Check if there is hearable audio sound from Speaker and AudioOutput port on rear panel.
  • Page 321 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-7-2 Apertur Lateral Min This test is used for detection of missing channels on Dachstein Probe Interface (DPI) or missing elements on probe, which are typically shown as dark Ultrasound Lines.
  • Page 322 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-7-3 Automatic Diagnostic Test. Automatic diagnostic test can be performed by selecting the desired test item from Selected Test item. 1.) Select the desired test item from "Selection Combo Box"...
  • Page 323 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-7-3-2 This diagnostic test performs the self-diagnosis of each vital product data for board or parts. 7-7-3-2-1 DRFM It verifies VPD information embedded in EEPROM on DRFM board and displays the part of VPD information and the test result as PASSED/FAILED.
  • Page 324 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-7-3-3 DRFM Overall Test This diagnostic test performs the self-diagnosis of each electric component and interface in the DRFM board. Table 7-2 below outlines Voluson™...
  • Page 325 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-7-3-4 This diagnostic test performs the self-diagnosis of every electric component and interface in DBM board. Table 7-3 below outlines Voluson™ P8 / Voluson™ P6 Diagnostics Tool described in the sub-sections. Table 7-3 Automatic diagnostic test under DBM Test Sub-section...
  • Page 326 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-7-3-4-8 CW(Continues Wave) It verifies signal path from AFE to DRFM by generating test wave signal and recording it with IQ phase recorder and displays each channels diagnostics result and the overall test result as PASSED/FAILED.
  • Page 327 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-7-3-6-5 DC4D Interface It verifies I2C path names I2C_DC4D by using EEPROM on DC4D board and displays the test result as PASSED/FAILED.
  • Page 328 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-7-3-9 HW Test Dialog This diagnostics test performs the self-diagnosis of each of manual function testing. 7-7-3-9-1 Aperture Lateral std It verifies system displays single aperture scan image on/off.
  • Page 329 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 7-8 Minimum Configuration to Boot/Scan 7-8-1 Booting without using the Ultrasound Hardware If the low-level ultrasound hardware is failing, the application can not be initialized correctly. Consequently, proper boot-up of the system is prevented.
  • Page 330 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4.) Set to “Disable” In case of Skylake Com Express, Move to “Advanced -> PCI Express Configuration -> PCI Express Port 0-3” 5.) Set to “Disable”...
  • Page 331 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) Press “F10” key and select “Yes” to save change. NOTE: How to check CPU code name, Ivybridge or Skylake. 1.) Turn the system on and go to BIOS mode by pressing “Del”...
  • Page 332 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Keep the H key pressed during complete startup (~ 1 minute). Following message appears: Figure 7-34 H Key pressed during startup 4.) Click on CONTINUE.
  • Page 333 7-9-1 Console Test 1.) Connect the USB "standard GE" service dongle on the front USB connector(s). As soon as the dongle is detected the login window pops up. 2.) Enter the "Standard GE Revolving" password (password changes every six months) and then click ENABLE.
  • Page 334 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 9.) The "TetrisConsole - Console Tester" window is displayed. Status Window Figure 7-37 Console Tester NOTE: For description of keys and their functions as well as the key code numbers refer to: Figure 5-6-2: OPIO layout on page 5-32.
  • Page 335 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 7-10 Troubleshooting Trees, Instructions and Tech Tips Table 7-8 below outlines Voluson™ P8 / Voluson™ P6 troubleshooting trees and instructions shown in the sub-sections.
  • Page 336 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-10-1 System does not boot up System doesn’t boot Verify that Is Standby Switch - Power Cord is connected to Mains - Power Switch is turned on illuminated? (on the backside of the system)
  • Page 337 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-10-2 Noise disturbs the Image Noise is shown in image. Is there any electrical device that affect the System ? Keep the System away from those electrical devices.
  • Page 338 ASIC ERVICE ANUAL 7-10-3 Trackball - Impaired Sensitivity The trackball has low sensitivity! Contact GE service representative Is it difficult to roll the to replace the trackball module. trackball? Set up the appropriate trackball response time (via [Trackball Speed] button in the System Setup - User Settings page).
  • Page 339 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-10-4 Printer Malfunction Figure 7-41 Printer Malfunction Chapter 7 - Diagnostics/Troubleshooting 7-47...
  • Page 340 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-10-5 Monitor Troubleshooting Fault symptom Check these items Check the HDMI and Power cable are properly connected to monitor. No image/No signal Check the HDMI and Power cable are properly connected to DRFM.
  • Page 341 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-10-6 DVD/CD+R/RW Drive Test NOTE: Only DVD+R/RW and CD+(R)W medias are available. 1.) Insert an empty DVD/CD+R/RW into the Drive. 2.) Enter “Patient Archive”...
  • Page 342 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-10-6 DVD/CD+R/RW Drive Test (cont?) 6.) Export images of the selected exam to DVD/CD+R/RW: a.) Click the EXPORT button. b.) If not already selected, choose “DVD/CD” drive from pop-up menu. c.) Enter “File name”.
  • Page 343 Try to ping from system to IP address on PC. If OK, hardware connection inside the system is OK. 7-10-7-2 GE remote service connection The customer gets visual information about GE remote service status (shown in the status area on the right bottom of the screen). Gray = Idle State Remote access is inactive.
  • Page 344 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-10-8 Tech Tips Table 7-9 below to access Voluson™ P8 / Voluson™ P6 (troubleshooting) Technical Tips listed. Each entry in the table will hyper-link to the issue, cause and solution. Table 7-9 Technical Tips Sub-section...
  • Page 345 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 7-11 Error Messages NOTICE If the problem (error message) still exists after performing the described actions, call technical support. !! NOTICE: Index Error Message...
  • Page 346 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Index Error Message Reason Action(s) 1. See error Restart the message. system 2. Stop scanning 1. See error message. Restart the 2.
  • Page 347 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL This page was intentionally left blank. Chapter 7 - Diagnostics/Troubleshooting 7-55...
  • Page 348 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7-56 Section 7-11 - Error Messages...
  • Page 349 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter contains replacement procedures for different modules and their subsystems. The Manpower, time and Tools indicated in the Sub-sections include all requirements from NOTICE Preparations to Installation Procedures.
  • Page 350 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 8-1 Chapter 8 Contents Section Description Page Number 8-14 8-34 Replacement of REAR CABINET BEZEL ASSY BT16 8-15 8-35 Replacement of ENCODER set 8-16 8-40...
  • Page 351 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 8-1 Chapter 8 Contents Section Description Page Number 8-49 8-151 Replacement of the DVD 8-50 8-153 Replacement of the OPIO USB Sub PWA 8-51 8-155 Replacement of AN KEY TOP...
  • Page 352 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-2 System Software - Installation/Upgrade Procedure Introduction There are two possibilities to update the system software: 1.) via the FMI from DVD button in the System Setup Service page Software parts to be upgraded (e.g., Ultrasound Application Software, Service Software, EUM, MS patches, etc.) and installation time depend on contents of the used System DVD.
  • Page 353 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-2-1-1 System Functional Checks Table 8-2 System Function Mode Task Expected Result Connect an abdominal probe, press the PROBE key, select the “Abdomen”...
  • Page 354 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-2-1-2 Check vital System Setup data 1.) Press the UTILITIES key on the control console. 2.) In the “Utilities” menu select SETUP button to invoke the setup desktop on the screen. 3.) On the left side of the screen select ADMINISTRATION and then click the SYSTEM INFO tab.
  • Page 355 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL NOTE: It is not necessary to note the Permanent Key; it is the state of the option that matters. Figure 8-2 System Setup - Administration - Option page 1.
  • Page 356 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-2-2 System Software - Installation Procedure (FMI from DVD) The system software installation procedure starts with saving and recording the settings present on the system (silent “Rollback”).
  • Page 357 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10.)Verify that only the DVD drive is connected to the system, then click OK (Figure 8-4 below). Figure 8-4 Verify that USB devices are disconnected, then click OK 11.)To start update procedure click YES The system saves Full Backup in silent mode on R:, then it reboots into LINUX.
  • Page 358 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 13.)Check disk is performed automatically - restart. Figure 8-7 Check disk is performed automatically 14.)3 dots (one after the other) appear on the screen (see: Figure 8-8 below).
  • Page 359 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 16.)Updating will take some time..Figure 8-9 Please wait... 17.)Please wait until all processes are finished (100 percent completed). 18.)The system is rebooting into windows (Boot screen - Voluson). NOTE: An automated process was developed to install the required software parts, perform check disk, remap driveletters and match settings.
  • Page 360 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 21.)If Control Panel is not working after first boot up, please shut down the system; then boot up again. NOTICE After turning off a system, wait at least 10 seconds before turning it on again.
  • Page 361 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-2-4 Software Update Package - Download/Installation Procedure 8-2-4-1 Overview This icon (at the right bottom of the screen) indicates that a new Software Update Package is available for download and installation.
  • Page 362 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) When downloading is finished, click Install. Figure 8-14 Download finished - click Install NOTE: Installation may take up to one hour. DO NOT interrupt the installation! If you want to install the new software later, click Exit.
  • Page 363 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8.) After clicking Install, the SW update procedure starts. The system saves Full Backup in silent mode on R:, then it reboots into LINUX. Figure 8-15 Installation in progress 9.) Check disk is performed automatically - restart.
  • Page 364 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 11.)Booting auto. Figure 8-17 Boot screen - auto 8-16 Section 8-2 - System Software - Installation/Upgrade Procedure...
  • Page 365 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 12.)Updating will take some time. First the image is saved to R:\ (Saving NTSF to image ..); then the image is restored from R:\ (Restoring NTFS from image ..). 13.)Please wait until all processes are finished (100 percent completed).
  • Page 366 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 14.)The system is rebooting into windows (Boot screen - Voluson). NOTE: An automated process was developed to install the required software parts, perform check disk, remap drive letters and match settings.
  • Page 367 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 15.)Wait until all processes are finished. Finally the 2D screen with the "New Software Verification" report is displayed on the monitor. Figure 8-19 New Software Verification 16.)Reconnect the external devices, install all the printers and adjust the printer settings as described Section 3-7 "Printer Installation"...
  • Page 368 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-3 Software and Functional Checks after Installation/Upgrade Procedure 1.) Press the UTILITIES key on the control panel. 2.) In the “Utilities” menu select SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select ADMINISTRATION and then click the OPTION tab.
  • Page 369 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 8-3 Software Upgrade Checklist Check Items Result The software was successfully installed without errors [___]Passed [___]Failed according to the instructions. Voluson SPC310 Software Part Number and Revision were correct as the instructions.
  • Page 370 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-4 Image Settings Only - Loading Procedure 8-4-1 Introduction The Image Settings contain: • Application Settings • 2D Factory and 2D user Presets •...
  • Page 371 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-7 Replacement or Activation of Options Following Software options are available: Table 8-4 Software options Options Option a) 3D/4D Activation Option a) Option b) Option b)
  • Page 372 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-7-1 Operation for activating Options 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select ADMINISTRATION and then click the OPTIONS tab (see:...
  • Page 373 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-8 Replacement of FRONT CABINET BEZEL ASSY BT16 8-8-1 Manpower 1 person, 1 minute 8-8-2 Tools Not required 8-8-3 Disassembly Procedure 1.) Remove the FRONT CABINET BEZEL ASSY BT16.
  • Page 374 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-9 Replacement of ASSY BEZEL FRONT DVD PLASTIC BT16 8-9-1 Manpower 1 people, 1 minute 8-9-2 Tools Not required 8-9-3 Disassembly Procedure 1.) Remove the ASSY BEZEL FRONT DVD PLASTIC BT16.
  • Page 375 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-10 Replacement of FRONT PRINTER BEZEL ASSY BT16 8-10-1 Manpower 1 people, 5 minutes 8-10-2 Tools Philips screw driver 8-10-3 Disassembly Procedure 1.) Remove the FRONT PRINTER BEZEL ASSY BT16.
  • Page 376 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-11 Replacement of Cover Top Cabinet Plastic 8-11-1 Manpower 1 people, 3 minutes 8-11-2 Tools Not required 8-11-3 Disassembly Procedure 1.) Remove the FRONT CABINET BEZEL ASSY BT16.
  • Page 377 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Remove the TOP CABINET COVER BT16. Refer to the figure below. TOP CABINET COVER BT16 Figure 8-27 Removing the TOP CABINET COVER BT16 8-11-4 Installation Procedure Install the new parts in the reverse order of removal procedure.
  • Page 378 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-12 Replacement of Front PEDESTAL COVER BT16 8-12-1 Manpower 1 people, 5 minutes 8-12-2 Tools Philips screw driver 8-12-3 Disassembly Procedure 1.) Remove the FRONT PRINTER BEZEL ASSY BT16.
  • Page 379 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Remove BOTTOM PEDESTAL COVER BT16. Refer to the figure below. BOTTOM PEDESTAL COVER BT16 Figure 8-30 Removing the BOTTOM PEDESTAL COVER BT16 4.) Unscrew 2 screws, and remove the Cover Front Pedestal.
  • Page 380 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-13 Replacement of TRAY PEDESTAL CABINET PLASTIC BT16 8-13-1 Manpower 1 people, 5 minutes 8-13-2 Tools Not required 8-13-3 Disassembly Procedure 1.) Remove the TRAY PEDESTAL CABINET PLASTIC BT16.
  • Page 381 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Install the Self-Adhesive Hook and Loop set on the TRAY PEDESTAL CABINET PLASTIC BT16. Refer to the figure below. Stick the self-adhesive hook and loop set on the TRAY PEDESTAL CABINET PLASTIC BT16.
  • Page 382 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-14 Replacement of REAR CABINET BEZEL ASSY BT16 8-14-1 Manpower 1 person, 1 minute 8-14-2 Tools Not required 8-14-3 Disassembly Procedure 1.) Remove the Rear Cabinet BEZEL ASSY BT16.
  • Page 383 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-15 Replacement of ENCODER set 8-15-1 Manpower One person, 40 minutes 8-15-2 Tools Philips screw driver, spanner 8-15-3 Removal procedure 1.) Unscrew 9 screws at the bottom of the keyboard body.
  • Page 384 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Lift up main keyboard assy. Refer to the figure below. Figure 8-39 Lifting up main keyboard assy 4.) Disconnect 4 cables. Refer to the figure below. Figure 8-40 Disconnecting 4 cables 8-36 Section 8-15 - Replacement of ENCODER set...
  • Page 385 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Pull 10 encoder key caps out. Refer to the figure below. Figure 8-41 Pulling 10 encoder key caps out 6.) Unscrew 2 screws on trackball and disconnect the cable to remove the trackball.
  • Page 386 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) After removing trackball, unscrew 21 screws and disconnect TGC cable. And lift up the main keyboard circuit board. Refer to the figure below. TCG cable Figure 8-43 Lifting up the main keyboard circuit board after unscrew 21 screws and disconnecting TGC cable 8-38...
  • Page 387 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8.) By using spanner, unscrew the nut of encoder to be replaced. And replace the encoder. Refer to the figures below. Figure 8-44 Unscrewing the nut of encoder to be replaced 9.) Pull the encoder out from the main keyboard circuit board.
  • Page 388 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-16 Replacement of Power ON/OFF key PCB Assembly 8-16-1 Manpower One person, 30 minutes 8-16-2 Tools Philips screw driver 8-16-3 Disassembly procedure 1.) Unscrew 9 screws at the bottom of the keyboard body.
  • Page 389 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Remove 5 probe holders. Refer to the figure below. Figure 8-47 Removing 5 probe holders 3.) Lift up main keyboard assy. Refer to the figure below. Figure 8-48 Lifting up main keyboard assy Chapter 8 - Replacement Procedures 8-41...
  • Page 390 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4.) Disconnect 4 cables under the main keyboard assy. Refer to the figure below. Figure 8-49 Disconnecting 4 cables 5.) Unscrew 2 screws on the system body. Refer to the figure below. Figure 8-50 Unscrewing 2 screws on the system body 8-42 Section 8-16 - Replacement of Power ON/OFF key PCB Assembly...
  • Page 391 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) Remove 2 RUBBER CAPS on SPEAKER COVER LEFT/RIGHT. Figure 8-51 Assembling 2 RUBBER CAPS on SPEAKER COVER LEFT/RIGHT 7.) Unscrew 2 screws and then remove the SPEAKER COVER LEFT/RIGHT. Refer to figure below. Figure 8-52 Unscrewing 2 screws to remove SPEAKER COVER LEFT/RIGHT Chapter 8 - Replacement Procedures 8-43...
  • Page 392 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8.) Unscrew 2 screws and remove the UPPER KEYBOARD ASSY. Figure 8-53 Removing Upper keyboard Assy 9.) Unscrew 7 screws then remove metal cover and USB connection circuit board. Refer to the figure below.
  • Page 393 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10.)Remove the On-Off key circuit. Refer to the figures below. Figure 8-55 Removing the On-Off key circuit 8-16-4 Installation procedure Install a new part in the reverse order of removal procedure.
  • Page 394 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-17 Replacement of Keyboard Assembly-TOP 8-17-1 Manpower One person, 25 minutes 8-17-2 Tools Philips screwdriver 8-17-3 Disassembly procedure 1.) Unscrew 9 screws at the bottom of the keyboard body. Refer to the figures below. Figure 8-56 Unscrew 9 screws 8-46 Section 8-17 - Replacement of Keyboard Assembly-TOP...
  • Page 395 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Remove 5 probe holders. Refer to the figure below. Figure 8-57 Removing 5 probe holders 3.) Lift up main keyboard assy. Refer to the figure below. Figure 8-58 Lifting up main keyboard assy Chapter 8 - Replacement Procedures 8-47...
  • Page 396 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4.) Disconnect 4 cables under the main keyboard assy. Refer to the figure below. Figure 8-59 Disconnecting 4 cables 5.) Unscrew 2 screws and remove cable tie on the system body. Refer to the figure below. Figure 8-60 Unscrewing 2 screws and removing cable tie on the system body 8-48 Section 8-17 - Replacement of Keyboard Assembly-TOP...
  • Page 397 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) Remove 2 RUBBER CAPS on SPEAKER COVER LEFT/RIGHT. Refer to the figure below. Figure 8-61 Removing 2 RUBBER CAPs on SPEAKER COVER LEFT/RIGHT 7.) Unscrew 2 screws and then remove the SPEAKER COVER LEFT/RIGHT.
  • Page 398 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8.) Unscrew 2 screws and pick up the sub keyboard. Refer to the figures below. Figure 8-63 Picking up the sub keyboard 8-17-4 Installation procedure Install a new part in the reverse order of removal procedure.
  • Page 399 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-18 Replacement of OPIO BUTTON SET 8-18-1 Manpower One person, 40 minutes 8-18-2 Tools Philips screwdriver 8-18-3 Removal Procedure 1.) Unscrew 9 screws at the bottom of the keyboard body.
  • Page 400 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Remove 5 probe holders. Refer to the figure below. Figure 8-65 Removing 5 probe holders 3.) Lift up main keyboard assy. Refer to the figure below. Figure 8-66 Lifting up main keyboard assy 8-52 Section 8-18 - Replacement of OPIO BUTTON SET...
  • Page 401 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4.) Disconnect 4 cables. Refer to the figure below. Figure 8-67 Disconnecting 4 cables Chapter 8 - Replacement Procedures 8-53...
  • Page 402 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Pull 10 encoder key caps out. Refer to the figure below. Figure 8-68 Pulling 10 encoder key caps out 6.) Unscrew 2 screws on trackball and disconnect the cable to remove the trackball.
  • Page 403 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) After removing trackball, unscrew 21 screws and disconnect TGC calbe. And lift up the main keyboard circuit board. Refer to the figure below. Figure 8-70 Lifting up the main keyboard circuit board after unscrewing 21 screws and disconnecting TGC cable 8.) Remove the button to be replaced.
  • Page 404 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-19 Replacement of OPIO Plastic lower cover assy 8-19-1 Manpower One person, 40 minutes 8-19-2 Tools Philips screwdriver 8-19-3 Disassembly procedure 1.) Unscrew 9 screws at the bottom of the keyboard body.
  • Page 405 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Remove 5 probe holders. Refer to the figure below. Figure 8-73 Removing 5 probe holders 3.) Lift up main keyboard assy. Refer to the figure below. Figure 8-74 Lifting up main keyboard assy Chapter 8 - Replacement Procedures 8-57...
  • Page 406 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4.) Disconnect 4 cables. Refer to the figure below. Figure 8-75 Disconnecting 4 cables 8-58 Section 8-19 - Replacement of OPIO Plastic lower cover assy...
  • Page 407 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Pull 10 encoder key caps and 8 TGC control key caps out. Refer to the figure below. Figure 8-76 Pulling 10 encoder key caps and 8 TGC control key caps out 6.) Disconnect A/N key cable and unscrew 10 screws of A/N key.
  • Page 408 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) Unscrew 4 screws of TGC control circuit board after disconnecting the cable. Refer to the figures below. Figure 8-78 Unscrewing 4 screws of TGC control circuit board 8.) Pull out the circuit board.
  • Page 409 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10.)Unscrew 21 screws from main keyboard. Then pull the main circuit board out. Refer to the figure below. Figure 8-81 Unscrewing 21 screws from main keyboard 8-19-4 Installation procedure Install a new part in the reverse order of removal procedure.
  • Page 410 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-20 Replacement of OPIO Plastic upper cover assy 8-20-1 Manpower One person, 30 minutes 8-20-2 Tools Philips screwdriver 8-20-3 Disassembly procedure 1.) Unscrew 9 screws at the bottom of the keyboard body.
  • Page 411 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Remove 5 probe holders. Refer to the figure below. Figure 8-83 Removing 5 probe holders 3.) Lift up main keyboard assy. Refer to the figure below. Figure 8-84 Lifting up main keyboard assy Chapter 8 - Replacement Procedures 8-63...
  • Page 412 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4.) Disconnect 4 cables under the main keyboard assy. Refer to the figure below. Figure 8-85 Disconnecting 4 cables 5.) Unscrew 2 screws on the system body. Refer to the figure below. Figure 8-86 Unscrewing 2 screws on the system body 8-64 Section 8-20 - Replacement of OPIO Plastic upper cover assy...
  • Page 413 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) Remove 2 RUBBER CAPS on SPEAKER COVER LEFT/RIGHT. Figure 8-87 Removing the RUBBER CAPS 7.) Unscrew 2 screws and then remove the SPEAKER COVER LEFT/RIGHT.. Refer to the figures below.
  • Page 414 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 9.) Unscrew 7 screws then remove metal cover, power On/Off key assembly and USB connection circuir board. Refer to the figure below. Figure 8-90 Unscrewing 7 screws 8-20-4 Installation procedure...
  • Page 415 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-21 Replacement of Monitor 8-21-1 Manpower 2 people, 10 minutes 8-21-2 Tools Philips screwdriver, Allen Wrench 8-21-3 Removal Procedure 1.) Unscrew 2 screws to remove the Monitor REAR CABINET BEZEL ASSY BT16.
  • Page 416 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Unscrew 5 screws to disconnect the LCD cables. Figure 8-92 Unscrewing 5 screws to disconnect the LCD cables 3.) One person grab the MONITOR. the other person should unscrew the 2 hexagonal socket screws and then separate the MONITOR for the LCD ARM.
  • Page 417 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-22 Replacement of the MONITOR ARM 8-22-1 Manpower Two persons, 20 minutes 8-22-2 Tools Philips screwdriver, Nipper, Allen Wrench 8-22-3 Removal Procedure 1.) Disassemble the LCD MONITOR.
  • Page 418 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4.) Unscrew 2 screws, and then remove SPEAKER COVER LEFT/RIGHT. Refer to the figure below. Figure 8-96 Unscrewing 2 screws to remove SPEAKER COVER LEFT/RIGHT 5.) Unscrew 1 screw to remove Monitor Cable Fix bracket.
  • Page 419 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) Unscrew LCD ARM STOPPER SCREW. Refer to figure below. Figure 8-99 Unscrewing LCD ARM STOPPER SCREW 8.) Pull out the LCD ARM from the LCD ARM HOLDER Figure 8-100 Pulling out the LCD ARM from the LCD ARM HOLDER 8-22-4 Installation Procedure...
  • Page 420 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-23 Replacement of LCD Cable 8-23-1 Manpower One person, 15 minutes 8-23-2 Tools Philips screw driver, Flat-blade screwdriver, wrench, Pincette, Cutter 8-23-3 Disassembly Procedure 1.) Unscrew 2 screws to remove the MONITOR REAR CABINET BEZEL ASSY BT16.
  • Page 421 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Unscrew 1 screw to remove MONITOR ARM COVER. Refer to the figure below. Figure 8-103 Unscrewing 1 screw to remove Monitor arm cover 4.) Unscrew 1 screw to remove MONITOR CABLE FIX BRKT.
  • Page 422 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Remove 2 RUBBER CAPS on SPEAKER COVER LEFT/RIGHT. Refer to the figure below. Figure 8-105 Removing 2 RUBBER CAPS on SPEAKER COVER LEFT/RIGHT 6.) Unscrew 2 screws, and then remove SPEAKER COVER LEFT/RIGHT.
  • Page 423 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8.) Lift up the OPIO and disconnect Power, OPIO, 2 USB cables from the OPIO HUB PWA. Refer to the figure below. Host interface 10 pin connector (OPIO cable) USB connector-1...
  • Page 424 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 9.) Remove the FRONT PRINTER BEZEL ASSY BT16. Refer to the figure below. FRONT PRINTER BEZEL ASSY BT16 Figure 8-109 Removing the FRONT PRINTER BEZEL ASSY BT16 TOP CABINET COVER BT16 Figure 8-110 Removing the TOP CABINET COVER BT16...
  • Page 425 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 11.)Unscrew 2 screws and remove FRONT PEDESTAL COVER BT16. FRONT PEDESTAL COVER BT16 Figure 8-112 Removing the FRONT PEDESTAL COVER BT16 12.)Untie the cable ties on keyboard base.
  • Page 426 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 14.)Unscrew 2 screws to remove POWER CORD FIX BRACKET. Refer to the figure below. Figure 8-115 Unscrewing 2 screws to remove POWER CORD FIX BRACKET 15.)Disassemble power cord and REAR CABINET BEZEL ASSY BT16.
  • Page 427 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 16.)Unscrew 7 screws and disassemble REAR EMI COVER. Figure 8-117 Unscrewing 7 screwing to disassemble REAR EMI COVER 17.)Pull out the LCD Cables through the hole. Refer to the figure below. Chapter 8 - Replacement Procedures 8-79...
  • Page 428 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 18.)Disconnect Monitor cables. Figure 8-118 Disconnecting Monitor cables 8-23-4 Assembly Procedure Install the new parts in the reverse order of disassembly procedure. 8-80 Section 8-23 - Replacement of LCD Cable...
  • Page 429 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-24 Replacement of ASSY Bumper Base Rubber 8-24-1 Manpower One person, 13 minutes 8-24-2 Tools Philips screwdriver 8-24-3 Disassembly Procedure 1.) Remove the FRONT CABINET BEZEL ASSY BT16.
  • Page 430 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Remove the TOP CABINET COVER BT16. Refer to the figure below. TOP CABINET COVER BT16 Figure 8-121 Remove the TOP CABINET COVER BT16 4.) Remove the REAR CABINET BEZEL ASSY BT16.
  • Page 431 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Unscrew 3 screws, and remove the RS Knobs. Refer to the figure below. Removing second RS Knob Removing fisrt RS knob Removing thrid RS Knob Figure 8-123 Removing the RS Knob 6.) Unscrew 4 screws, and remove the TRAY SIDE CABINET BT16.
  • Page 432 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) Unscrew 2 screws, and remove the ASSY Left Side plastic Cover. Refer to the figure below. Figure 8-125 Removing the ASSY Left Side Plastic Cover 8.) Unscrew 2 screws, and remove the ASSY RIGHT SIDE PLASTIC COVER BT16.
  • Page 433 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 9.) Remove the COVER CASTOR NARROW L/R Plastic. Refer to the figure below. Figure 8-127 Removing the COVER CASTOR NARROW L/R Plastic Chapter 8 - Replacement Procedures 8-85...
  • Page 434 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10.)Remove the Bumper Base Rubber. Refer to the figure below. Figure 8-128 Removing the Bumper Base Rubber 8-24-4 Installation Procedure Install the new parts in the reverse order of removal procedure.
  • Page 435 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-25 Replacement of OPIO 8-25-1 Manpower One person, 15 minutes 8-25-2 Tools Philips screwdriver, Nipper, Allen Wrench 8-25-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 436 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Remove the 5 probe holders. Refer to the figure below. Figure 8-131 Removing the 5 probe holders 4.) Unscrew 9 screws located in the bottom of the OPIO. Refer to the figure below. Figure 8-132 Unscrewing 9 screws located on the bottom of the OPIO 8-88 Section 8-25 - Replacement of OPIO...
  • Page 437 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Lift up the OPIO and disconnect Power, OPIO, 2 USB cables from the OPIO HUB PWA. Refer to the figure below. Host interface 10 pin connector (OPIO cable) USB connector-1...
  • Page 438 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) Unscrew 4 screws to remove upper keyboard assembly. Figure 8-135 Unscrewing 4 screws to remove upper keyboard assembly 8-25-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
  • Page 439 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-26 Replacement of the SPEAKER 8-26-1 Manpower One person, 30 minutes 8-26-2 Tools Philips screwdriver 8-26-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 440 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-27 Remove the PEDESTAL BOTTOM COVER BT16 8-27-1 Manpower One person, 2 minutes 8-27-2 Tools 8-27-3 Preparations 8-27-4 Removal Procedure 1.) Remove PEDESTAL BOTTOM COVER BT16.
  • Page 441 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-28 Replacement of the NEST 8-28-1 Manpower One person, 10 minutes 8-28-2 Tools Philips screwdriver, Allen Wrench 8-28-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 442 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Remove the REAR CABINET BEZEL ASSY BT16. Refer to the figure below. REAR CABINET BEZEL ASSY BT16 Figure 8-140 Removing the Rear Cabinet Bezel Assy BT16 3.) Unscrew 7 screws and remove the REAR EMI COVER BRKT.
  • Page 443 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4.) Disconnect all connectors. Figure 8-142 Disconnection all cables Chapter 8 - Replacement Procedures 8-95...
  • Page 444 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Set the 3 RS KNOBS to unlock position to separate from the SIDE COVER R. and unscrew one screw from the each 3 RS KNOB. Refer to the figure below. Removing first RS Knob Removing second RS Knob Removing third RS Knob...
  • Page 445 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) Unscrew 4 screws and remove the Nest box. Refer to the figure below. Figure 8-144 Unscrewing 4 screws and removing the NEST BOX 8-28-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
  • Page 446 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-29 Replacement of the AIR FILTER 8-29-1 Manpower One person, 1 minutes 8-29-2 Tools None 8-29-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 447 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-29-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. Chapter 8 - Replacement Procedures 8-99...
  • Page 448 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-30 Replacement of the Probe Holder(KIT) 8-30-1 Manpower One person, 1 minutes 8-30-2 Tools None 8-30-3 Removal Procedure 1.) Remove 5 probe holders. Refer to the figure below. Figure 8-147 Removing the 5 PROBE HOLDERs 8-30-4 Installation Procedure...
  • Page 449 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-31 Replacement of SIDE COVER L ASSY BT16 8-31-1 Manpower One person, 5 minutes 8-31-2 Tools Philips Screwdriver 8-31-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page...
  • Page 450 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Remove the TOP CABINET COVER BT16. Refer to the figure below. TOP CABINET COVER BT16 Figure 8-150 Removing the TOP CABINET COVER BT16 4.) Unscrew 2 screws, and remove the ASSY LEFT SIDE PLASTIC COVER BT16.
  • Page 451 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-32 Replacement of SIDE COVER R ASSY 8-32-1 Manpower One person, 5 minutes 8-32-2 Tools Philips Screwdriver 8-32-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 452 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Remove the COVER TOP CABINET PLASTIC BT16. Refer to the figure below. COVER TOP CABINET PLASTIC BT16 Figure 8-154 Removing the COVER TOP CABINET PLASTIC BT16 4.) Unscrew 3 screws, and remove the RS Knobs.
  • Page 453 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Unscrew 2 screws, and remove the ASSY RIGHT SIDE PLASTIC COVER BT16. Refer to the figure below. ASSY RIGHT SIDE PLASTIC COVER BT16 Figure 8-156 Removing the ASSY RIGHT SIDE PLASTIC COVER BT16 8-32-5...
  • Page 454 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-33 Replacement of the Trackball 8-33-1 Manpower One person, 20 minutes 8-33-2 Tools Philips Screwdriver 8-33-3 Removal procedure 1.) Unscrew 9 screws at the bottom of the keyboard body. Refer to the figures below. Figure 8-157 Unscrewing 9 screws 8-106 Section 8-33 - Replacement of the Trackball...
  • Page 455 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Remove 5 probe holders. Refer to the figure below. Figure 8-158 Removing 5 probe holders 3.) Lift up main keyboard assy. Refer to the figure below. Figure 8-159 Lifting up main keyboard assy Chapter 8 - Replacement Procedures 8-107...
  • Page 456 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4.) Disconnect 4 cables. Refer to the figure below. Figure 8-160 Disconnecting 4 cables 8-108 Section 8-33 - Replacement of the Trackball...
  • Page 457 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Unscrew 2 screws on trackball and disconnect the cable side of the trackball. Refer to the figures below. Figure 8-161 Unscrewing 2 screws on trackball 8-33-4 Installation Procedure 1.) Put the trackball guide ring.
  • Page 458 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Connect the cable to the connector on the main PWA and screw the trackball with 2 screws. Refer to the figure below. Figure 8-163 Connecting the cable to the connector and screwing the trackball NOTE: Please check the position of the label as the below figure to connect in the right position.
  • Page 459 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-34 Replacement of the TGC BOARD 8-34-1 Manpower One person, 10 minutes 8-34-2 Tools Philips Screwdriver 8-34-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 460 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4.) Unscrew 4 screws, then pull out the TGC BOARD. Refer to the figure below. Figure 8-165 Unscrewing 4 screws and pulling out the TGC BOARD 8-34-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
  • Page 461 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-35 Replacement of AN KEYBOARD ASSY 8-35-1 Manpower One person, 15 minutes 8-35-2 Tools Philips Screwdriver 8-35-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 462 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-35-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-20: Replacement of AN KEYBOARD ASSY - Functional Tests. 8-114 Section 8-35 - Replacement of AN KEYBOARD ASSY...
  • Page 463 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-36 Replacement of the Caster Cap BT16 8-36-1 Manpower One person, 2 minutes 8-36-2 Tools None 8-36-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 464 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Remove the FRONT PRINTER BEZEL ASSY BT16. Refer to the figure below. FRONT PRINTER BEZEL ASSY BT16 Figure 8-169 Removing the FRONT PRINTER BEZEL ASSY BT16 3.) Remove the Top Cabinet Cover BT16.
  • Page 465 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4.) Remove the Rear Cabinet Bezel Assy BT16. Refer to the figure below.v Rear Cabinet Bezel Assy BT16 Figure 8-171 Removing the Rear Cabinet Bezel Assy BT16 Chapter 8 - Replacement Procedures 8-117...
  • Page 466 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Unscrew 3 screws, and remove the RS Knobs. Refer to the figure below. Remove second RS Knob Remove first RS Knob Remove third RS Knob Figure 8-172 Removing the RS Knob 6.) Unscrew 4 screws, and remove the SIDE CABINET TRAY BT16.
  • Page 467 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) Unscrew 2 screws, and remove the SIDE COVER L ASSY BT16 . Refer to the figure below. SIDE COVER L ASSY BT16 Figure 8-174 Removing the SIDE COVER L ASSYBT16 8.) Unscrew 2 screws, and remove the SIDE COVER R ASSYBT16 .
  • Page 468 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 9.) Remove the COVER CASTOR NARROW L/R BT16. Refer to the figure below. COVER CASTOR NARROW L/R BT16 Figure 8-176 Removing the COVER CASTOR NARROW L/R BT16 8-36-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
  • Page 469 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-37 Replacement of the CASTER 8-37-1 Manpower One person, 10 minutes 8-37-2 Tools Monkey Wrench 8-37-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 470 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Figure 8-178 Unscrewing the nut 8-37-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 8-122 Section 8-37 - Replacement of the CASTER...
  • Page 471 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-38 Replacement of the SIDE CABINET TRAY BT16 8-38-1 Manpower One person, 2 minutes 8-38-2 Tools Philips screwdriver 8-38-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page...
  • Page 472 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-39 Replacement of the REAR HANDLE 8-39-1 Manpower One person, 5 minutes 8-39-2 Tools Philips screwdriver, Allen Wrench 8-39-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page...
  • Page 473 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Remove the TOP CABINET COVER BT16. Refer to the figure below. TOP CABINET COVER BT16 Figure 8-181 Removing the TOP CABINET COVER BT16 3.) Remove BOTTOM PEDESTAL COVER BT16.
  • Page 474 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Unscrew 2 screws, and remove the REAR HANDLE. Refer to the figure below. Figure 8-184 Removing the REAR HANDLE 8-39-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
  • Page 475 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-40 Replacement of the SPEAKER COVER LEFT/RIGHT 8-40-1 Manpower One person, 5 minutes 8-40-2 Tools Philips screwdriver 8-40-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 476 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-41 Replacement of the DPS/DPS with Battery 8-41-1 Manpower One person, 30 minutes 8-41-2 Tools Philips screwdriver, Allen Wrench 8-41-3 Preparations 1.) Power Off/Shutdown the system as described in...
  • Page 477 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-41-4 Removal Procedure Figure 8-187 DPS 1.) Draw the DPS out of the Nest Frame. Figure 8-188 Drawing the DPS 8-41-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
  • Page 478 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-42 Replacement of the DRFM 8-42-1 Manpower One person, 30 minutes 8-42-2 Tools Philips screwdriver, Allen Wrench 8-42-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 479 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-42-4 Removal Procedure DRFM Figure 8-189 DRFM 1.) Draw the DRFM out of the Nest Frame. Figure 8-190 Drawing the DRFM Chapter 8 - Replacement Procedures 8-131...
  • Page 480 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-42-5 Installation Procedure 8-42-5-1 DRFM (5497712) Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perfom 8-42-6 Serial Number - Reprogamming Procedure to reprogram the serial number.
  • Page 481 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-42-6 Serial Number - Reprogamming Procedure NOTICE To reprogramming the System Serial Number, contact the OLC(On Line Center) to get the necessary "Serial Key"...
  • Page 482 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Hardware ID Figure 8-194 Hardware ID on label DRFM 2.) Call the OLC to get the "Serial Key" and the "Permanent Key" 3.) Program the System Serial number and re-enter the Permanent Key To program the System SNO, click the OPTIONS tab: Field to enter...
  • Page 483 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL a.) Enter the "Serial key" (alphanumeric letters based on system serial number and DRFM Hardware ID) in the field "Serial Number" (e.g., AEE1193428AA6078123456SU7). Then press ENTER.
  • Page 484 Software patch does not need to be installed from software version Ext.5 or later. NOTE: Software patch can also be downloaded at: http://ge.box.com/s/yspo13ahrk3uq79gmif10api72p73f1 1.) Insert Software patch CD on the system. 2.) Click the ASI button for patch (Util - System setup - Administration - Service tab - Service tool).
  • Page 485 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Choose media to DVD/CD and click Load. Figure 8-197 Choosing media to DVD/CD 4.) Click ‘OK’. Figure 8-198 Click OK 5.) Insert ‘y’...
  • Page 486 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) If “Operation was a success” is displayed on command window, patch is well installed. Figure 8-199 “Operation was a success” is displayed on command window 7.) Shutdown to replace DRFM.
  • Page 487 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-43 Replacement of the SOM in DRFM 8-43-1 Manpower One person, 30 minutes 8-43-2 Tools Philips screwdriver 8-43-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 488 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL When replacing with new SOM, Push the SOM in both side of DRFM ASSEMBLY without CAUTION putting weight on the board. !! CAUTION: 8-43-5 Installation Procedure...
  • Page 489 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-44 Replacement of the HDD in DRFM 8-44-1 Manpower One person, 30 minutes 8-44-2 Tools Philips screwdriver 8-44-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 490 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-45 Replacement of the DBM64 8-45-1 Manpower One person, 30 minutes 8-45-2 Tools Philips screwdriver 8-45-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 491 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Unscrew 6 screws. 3.) Remove DBM64 from the EMI SHIELD BRACKET. EMI SHIELD BRACKET Figure 8-204 Unscrewing 6 screws and removing the EMI SHIELD BRACKET 8-45-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
  • Page 492 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-46 Replacement of the T-DPI 8-46-1 Manpower One person, 30 minutes 8-46-2 Tools Philips screwdriver, Allen Wrench, Hirose ejector 8-46-3 Preparations 1.) Power Off/Shutdown the system as described in...
  • Page 493 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-46-4 Removal Procedure T-DPI Figure 8-205 DBM/T-DPI 1.) Draw the DBM/T-DPI out of the Nest Frame. Figure 8-206 Drawing the DBM/T-DPI 2.) Unscrew 6 screws.
  • Page 494 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Remove T-DPI from the EMI SHIELD BRACKET. Figure 8-207 Removing the EMI SHIELD BRACKET Figure 8-208 Bottom of T-DPI 8-46-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
  • Page 495 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-47 Replacement of the CSI 8-47-1 Manpower One person, 30 minutes 8-47-2 Tools Philips screwdriver 8-47-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 496 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Unscrew 4 screws to remove CSI board and Separate the LED connector. HDD Status Network Status Network Speed Orange: Megabit Figure 8-210 Seperating LED connector/FAN connector 8-47-5 Installation Procedure...
  • Page 497 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-48 Replacement of the CBP 8-48-1 Manpower One person, 30 minutes 8-48-2 Tools Philips screwdriver, Allen Wrench 8-48-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 498 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Unscrew 6 screws. Figure 8-212 Unscrewing 6 screws 8-48-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-39: Replacement of the CBP - Functional Tests.
  • Page 499 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-49 Replacement of the DVD 8-49-1 Manpower One person, 10 minutes 8-49-2 Tools Philips screwdriver 8-49-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 500 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Unscrewing 2 screws to remove the DVD . Figure 8-215 Removing the DVD 4.) Unscrewing 4 screws to remove the SIDE STRIPS. Figure 8-216 Removing the SIDE STRIPS 8-49-5 Installation Procedure...
  • Page 501 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-50 Replacement of the OPIO USB Sub PWA 8-50-1 Manpower One person, 30 minutes 8-50-2 Tools Philips screwdriver 8-50-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page...
  • Page 502 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Unscrew 5 screws and disconnect 2 USB cables, Then Remove 2 Brackets and OPIO USB Sub PWA. Figure 8-218 Removing 2 Brackets and OPIO USB Sub PWA 8-50-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
  • Page 503 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-51 Replacement of AN KEY TOP 8-51-1 Manpower One person, 30 minutes 8-51-2 Tools Flat head screw driver or Flat tool 8-51-3 Preparations 1.) Power Off/Shutdown the system as described in...
  • Page 504 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-52 Replacing optional Peripherals NOTICE Normally auxiliary devices and peripherals come pre-installed with the Voluson™ P8 / Voluson™ P6 system.
  • Page 505 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-52-3 Printer Installation (BW and Color) 8-52-3-1 BW Printer: SONY UP-D898, Color Printer: SONY UP-D25MD 1.) Assemble 2 EMI gaskets (5448934 - EMI gasket, side, b-w printer bkt-3) on printer base plate. 2.) Assemble 2 EMI gaskets (5447303 - EMI gasket, side, b-w printer bkt-2) on printer base plate.
  • Page 506 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Disassemble FRONT PRINTER BEZEL ASSY BT16. 6.) Remove PRINTER DUMMY COVER PLATE BT16. Figure 8-222 Disassembling PRINTER DUMMY COVER PLATE BT16 7.) Unscrew 2 screws to disassemble power cord fix bracket.
  • Page 507 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 9.) Unscrew 7 screws and disassemble REAR EMI COVER. 10.)Unscrew 2 screws and disassemble printer EMI cover 1 and 2. Figure 8-224 Disassembling printer EMI covers Chapter 8 - Replacement Procedures 8-159...
  • Page 508 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 11.)Assemble Printer assembly. 12.)Push the printer assembly till it makes contact with sheet metal chassis. Push the printer assembly Printer Assembly Figure 8-225 Assembling printer 13.)Carefully insert the printer base plate in to the chassis slot.
  • Page 509 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 16.)Insert the printer USB cable and printer power cable. 17.)Insert the printer power cable in to the nest box. 18.)Insert the printer USB cable in to the nest box. Inserting Printer USB cable Printer Printer...
  • Page 510 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 21.)Screw the 7 screws and fix the REAR EMI COVER. 22.)Assemble the REAR CABINET BEZEL ASSY BT16, power cord and POWER CORD FIX BRACKET.
  • Page 511 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-52-3-2 Color Printer Assembly 1.) Assemble 2 EMI gaskets (5448934 - EMI gasket, side, b-w printer bkt-3) on printer base plate. 2.) Assemble 2 EMI gaskets (5448934 - EMI gasket, side, b-w printer bkt-3) on printer base plate.
  • Page 512 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 5.) Remove the FRONT PRINTER BEZEL ASSY BT16. Figure 8-231 FRONT PRINTER BEZEL ASSY BT16 6.) Disassemble FRONT PRINTER BEZEL ASSY BT16. POWER CORD FIX BRACKET REAR CABINET BEZEL ASSY BT16 Figure 8-232 Removing POWER CORD FIX BRACKET and REAR CABINET BEZEL ASSY BT16...
  • Page 513 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) Unscrew 7 screws and disassemble REAR EMI COVER. 8.) Unscrew 2 screws and disassemble printer EMI cover 1 and 2. 2 screwsscrews and printer EMI Unscrew 7 screws Rear side view after removing the...
  • Page 514 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 13.)Assemble the printer assembly. 14.)Assemble the printer base plate with chassis using 2 screws (2159634 - screw BH M4X10). 2 screws (2159634 -screw BH M4X10) Figure 8-235 Assembling printer base plate 15.)Insert the printer USB cable and printer power cable.
  • Page 515 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 17.)Insert the printer USB cable in to the nest box. 18.)Route the color printer USB and power cable through the left side cable clamp. Left side cable clamp Printer USB cable in Nest box Figure 8-237 Routing the cables...
  • Page 516 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 21.)Assemble the DVD front cover plastic with DVD strip and B/W printer strip using 8 screws (2327596 - screw washer head M3x6). 22.)Assemble 4 numbers of Holding Catch, female with DVD front cover using 8 screws (2306562 - screw FH M3x6) Figure 8-239 Assembling DVD front cover...
  • Page 517 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 25.)Assemble the FRONT CABINET BEZEL ASSY BT16. Figure 8-241 Assembling the FRONT CABINET BEZEL ASSY BT16 Chapter 8 - Replacement Procedures 8-169...
  • Page 518 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-52-4 HORIZONTAL TV PROBE HOLDER Installation 1.) Install HORIZONTAL TV PROBE HOLDER on the left side bottom of OPIO with 3 screws. HORIZONTAL TV PROBE HOLDER Figure 8-242 Assembling HORIZONTAL TV PROBE HOLDER with 3 screws Figure 8-243 HORIZONTAL TV PROBE HOLDER is installed on Left Side only...
  • Page 519 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-52-5 VERTICAL TV PROBE HOLDER Installation 1.) Install VERTICAL TV PROBE HOLDER BASE on the right side bottom of OPIO with 2 screws. Figure 8-244 Assembling TV PROBE HOLDER BASE with 2 screws (2159634, BH M4x10) 2.) Place the TV PROBE HOLDER RUBBER on the PROBE HOLDER BASE.
  • Page 520 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-53 Replacement of FAN with Cable ASSY II 8-53-1 Manpower One person, 30 minutes. 8-53-2 Tools Philips screwdriver, Allen Wrench 8-53-3 Preparations 1.) Power Off/Shutdown the system as described in...
  • Page 521 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-54 Functional Checks after FRU replacement (Debrief Guidelines) 1.) Perform required Functional and/or Leakage Current tests based upon the FRU being replaced. 2.) Clarify the scripts to debrief a Service Dispatch using your pole's dispatch tool.
  • Page 522 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-55 Replacement of UVC Replace FRU part as described in system's UVC Installation Kit Installation Instruction (5758249) in CDL. 8-174 Section 8-55 - Replacement of UVC...
  • Page 523 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-56 Replacement of CWD piggy 8-56-1 Manpower One person, 30 minutes. 8-56-2 Tools Philips screwdriver, Hirose ejector 8-56-3 Preparations Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 524 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 2.) Remove REAR CABINET BEZEL ASSY BT16 Figure 8-248 REAR CABINET BEZEL ASSY BT16 3.) Unscrew 7 screws and disassemble the REAR EMI COVER. Figure 8-249 Unscrewing 7 fastners and disassembling the REAR EMI COVER 8-176 Section 8-56 - Replacement of CWD piggy...
  • Page 525 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 4.) Unplug all cables and Unscrew 7 screws and remove the NEST EMI BRKT BT16. Figure 8-250 Unpluging all cables and removing the NEST EMI BRKT BT16 5.) Remove 3 RS KNOBS .
  • Page 526 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) Pull out DBM64/T-DPI Boards. Figure 8-252 Pulling out DBM64/T-DPI Boards 8-178 Section 8-56 - Replacement of CWD piggy...
  • Page 527 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) Unscrew 1 screw and replace the CWD Piggy. Figure 8-253 Unscrewing 1 screw and replacing the CWD piggy 8-56-5 Installation Procedure 1.) Install the new board assy in the reverse order of removal procedure.
  • Page 528 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-57 Replacement of ECG Kit assy 8-57-1 Manpower One person, 10 minutes. 8-57-2 Tools Philips screwdriver 8-57-3 Preparations Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 529 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Open keyboard assembly. 4.) Remove ECG DUMMY BRKT. Figure 8-255 ECG DUMMY BRKT 8-57-5 Installation Procedure 1.) Assemble ECG kit on keyboard base. 2.) Screw ECG kit with Keyboard base using 2 screws (2159625-screw w/SP PH m4x8)and fix ECG USB cable by using the hook of bracket.
  • Page 530 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Connect the USB cable with Keyboard assembly as shown. 4.) Assemble Keyboard assembly. Figure 8-257 USB cable 5.) Screw 9 numbers of screws below the . 6.) Assemble 5 probe holders.
  • Page 531 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 7.) Attach ECG label. ECG label Figure 8-259 ECG label 8.) Perform: FRU8-49: Replacement of ECG Kit assy - Functional Tests Chapter 8 - Replacement Procedures 8-183...
  • Page 532 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-58 Replacement of Battery Pack Assy 8-58-1 Manpower One person, 20 minutes 8-58-2 Tools Philips screwdriver 8-58-3 Preparation Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
  • Page 533 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Disconnect Battery power and signal cables. Figure 8-261 Disassembling Battery power and signal cables from DPS with Battery board 4.) Open cover and replace Battery pack from DPS with Battery top cover.
  • Page 534 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 8-59 Replacement of WIRELESS LAN 8-59-1 Manpower One person, 30 minutes 8-59-2 Tools Standard Field Service Tool Kit 8-59-3 Preparation 1.) Power Off/Shutdown the system as described in 4-3-2 on page...
  • Page 535 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 3.) Diconnect the WERELESS LAN cable from NEST BOX. Figure 8-265 Disconnecting WIRELESS LAN cable from NEST BOX 4.) Untie the cable clamp. Figure 8-266 Untying cable clamp to unroute WIRELESS LAN Cable 5.) Pull out WIRELESS LAN dongle from loop sheet.
  • Page 536 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 6.) Unfold WIRELESS LAN and Check the dongle label before installation. WIRELESS LAN dongle Figure 8-268 Checking the dongle label 7.) Attach loop sheet in WIRELESS LAN dongle.
  • Page 537 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 8-59-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform : 3.) Perform : 8-59-6 Set up WIRELESS LAN 1.) Press Utility ->...
  • Page 538 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL FRU8-2: System Software - Installation/Upgrade Procedure Functional and/or See Section Leakage Current Test Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, Voluson™...
  • Page 539 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL FRU8-8: Replacement of Monitor Functional and/or See Section Leakage Current Test Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, Voluson™...
  • Page 540 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL FRU8-13: Replacement of the NEST Functional and/or See Section Leakage Current Test Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, Voluson™...
  • Page 541 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL FRU8-28: Replacement of the DPS/DPS with Battery Functional and/or See Section Leakage Current Test Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, Voluson™...
  • Page 542 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL FRU8-36: Replacement of the DBM64 Functional and/or See Section Leakage Current Test Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, Voluson™...
  • Page 543 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL FRU8-40: Replacement of the DVD Functional and/or See Section Leakage Current Test Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, Voluson™...
  • Page 544 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL FRU8-44: Replacing optional Peripherals Functional and/or See Section Leakage Current Test Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, Voluson™...
  • Page 545 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL FRU8-49: Replacement of ECG Kit assy Functional and/or See Section Leakage Current Test Debrief Script 10-8-5 Enclosure Leakage Current Test Proprietary Service Manual, Voluson™...
  • Page 546 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 8-198 Section 8-59 - Replacement of WIRELESS LAN...
  • Page 547 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives an overview of replacement parts available for the Voluson™ P8 / Voluson™ P6. Table 9-1 Contents in Chapter 9 Section...
  • Page 548 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-2 List of Abbreviations Alternating Current Analog to Digital Converter A/N Key Alpha Numeric Keys ASIC Application Specific Integrated Circuit Assy Assembly DBM64...
  • Page 549 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-3 Parts List Groups Table 9-2 Mechanical and user accessible parts Item Part Group Name Table Number Description Table 9-3 on page 9-5 Housing Covers 101-120 Plastic Covers...
  • Page 550 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-4 Plastics Covers (Front/Sides/Rear) Figure 9-1 Plastic Covers Section 9-4 - Plastics Covers (Front/Sides/Rear)
  • Page 551 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 9-3 Plastics Covers Part Item Part Name Number Description BT13 BT14 BT15 BT16 COVER CASTER 5728121 COVER CASTER NARROW SET BT16 NARROW SET BT16 Left side Plastic cover Voluson...
  • Page 552 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-5 LCD Monitor Figure 9-2 LCD Monitor Table 9-4 LCD Monitor Item Part Name Part Number Description BT13 BT14 BT15 BT16 5669733...
  • Page 553 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-6 OPIO 311~322 Figure 9-3 OPIO Chapter 9 - Renewal Parts...
  • Page 554 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 9-5 OPIO Part Item Part Name Number Description BT13 BT14 BT15 BT16 VP8P6 keyboard Assembly- VP8P6 keyboard Assembly- 5451539-2 TOP BT16 TOP BT16 VP8P6Lower OPIO...
  • Page 555 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 9-5 OPIO Part Item Part Name Number Description BT13 BT14 BT15 BT16 AN Key Top with Base AN Key Top with Base 5722479 Assembly, FinnishVP6_P8 Assembly, FinnishVP6_P8...
  • Page 556 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-7 Nest Box Parts Figure 9-4 Nest Box parts 9-10 Section 9-7 - Nest Box Parts...
  • Page 557 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 9-6 Nest Box Part Item Part Name Number Description BT13 BT14 BT15 BT16 T-DPI PWA 5719539-2 Probe Interface Board Assy assy 5573639-2 64ch Dachstein Beamformer Mla4...
  • Page 558 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 9-6 Nest Box Part Item Part Name Number Description BT13 BT14 BT15 BT16 Manganese Dioxide Lithum Coin Battery, Manganese 5173038 CR2450 Dioxide Lithum...
  • Page 559 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-8 Mechanical Parts Figure 9-5 Mechanical Parts Chapter 9 - Renewal Parts 9-13...
  • Page 560 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 9-7 Mechanical Parts Part Item Part Name Number Description BT13 BT14 BT15 BT16 5400391-2 Total Lock Caster Total Lock Caster 5400392-2 Caster with swivel available Swivel Lock Caster...
  • Page 561 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-9 Options Figure 9-6 Options Table 9-8 Options Part Item Part Name Number Description BT13 BT14 BT15 BT16 5408092 Vertical TV probe holder Vertical TV probe holder Horizontal TV probe...
  • Page 562 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-10 Optional Peripherals and Accessories 9-10-1 Printers and options Figure 9-7 Optional Peripherals and Accessories - Printers 9-16 Section 9-10 - Optional Peripherals and Accessories...
  • Page 563 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 9-9 Optional Peripherals and Accessories Part Item Part Name Number Description BT13 BT14 BT15 BT16 5692801-2 Voluson sysDVD Software SPC310 Ext.4 5692801-3 Voluson sysDVD Software SPC310 Ext.5 Voluson P6/P8...
  • Page 564 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-11 System Manuals Table 9-10 System Manuals - Voluson™ P8 / Voluson™ P6 Part Name Part # VP8/VP6 5723243 Service Manual System User Manuals 5654430-100...
  • Page 565 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-12 Probes Table 9-11 below outlines the replacement parts described in the sub-sections. Table 9-11 Probes - Replacement Parts Sub-section Description Page Number...
  • Page 566 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 9-12-2 2D-Probes - Linear Array Transducer Figure 9-9 2D Linear Array Transducer Table 9-13 2D linear array Transducers Item Part Name Part Number Description Small and lightweight linear-array transducer 4.0 - 10.0 MHz, 192...
  • Page 567 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 9-12-3 Real-Time 4D Volume Probes Figure 9-10 Real-Time 4D Volume Probes Table 9-14 Real-Time 4D Volume Probes Item Part Name Part Number Description Real-time 4D endocavity broadband electronic curved...
  • Page 568 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 9-12-4 2D-Probes - Phased Array Transducers Figure 9-11 2D Phased Array Probe Table 9-15 2D-Probes - Phased Array Transducers Item Part Name Part Number Description...
  • Page 569 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 9-12-5 Probe-Compatibility Table 9-16 Probe Compatibility for Voluson P Series Affected Products Part Information BT16 BT13 BT14 BT15 (SPC310) Array type Part Number Part Description VP6 VP8 VP6 VP8 VP6 VP8 VP6 VP8 5488477...
  • Page 570 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-13 Power Cord Figure 9-12 Power Cords 9-24 Section 9-13 - Power Cord...
  • Page 571 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 9-17 Power Cord Item Part Name Part Number Description Power Cord for India 5182611 Power Cord for India Power Cord for Switzerland 5182235 Power Cord for Switzerland Power Cord for UK/HK...
  • Page 572 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 9-14 Biopsy Needle Guide 1001 1002 1004 1003 1005 1006 1007 1009 1008 Figure 9-13 Biopsy Needle Guides Table 9-18 Biopsy Needle Guides Item Part Name...
  • Page 573 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 9-18 Biopsy Needle Guides Item Part Name Part Number Description Non Sterile Multi Angle Bracket needle guide for probe 3Sc-RS 3Sc-RS H46222LC 1005...
  • Page 574 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 9-28 Section 9-14 - Biopsy Needle Guide...
  • Page 575 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your Voluson™ P8 / Voluson™ P6 system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
  • Page 576 Your GE Service Representative has an in-depth knowlegde of your Voluson™ P8 / Voluson™ P6 ultrasound scanning system and can best provide competent, efficient service.
  • Page 577 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10-3-1 How often should care & maintenance tasks be performed? (cont?) Table 10-2 Customer Care Schedule Item Service at Indicated Time Notes •...
  • Page 578 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 10-2 Customer Care Schedule Item Service at Indicated Time Notes Endocavity Probe Also after corrective • Leakage maintenance or as required by Current Checks your facilities QA program.
  • Page 579 OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 10-4 Tools Required Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID Description Tool ID Description 9-45358 Pliers Retaining Ring 9-XL9971MM Xcelite-hex Blade 1.27mm 9-4078 Scribe 9-XL9972MM Xcelite-hex Blade 1.5mm...
  • Page 580 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-46696 Hex Keys 9-XL100X Wire Stripper/Cutter 5 inch - 100X 9-39829 Torpedo Level, Magnetic...
  • Page 581 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 10-5 Tools Required 10-5-1 Special Tools, Supplies and Equipment 10-5-1-1 Specific Requirements for Care & Maintenance Table 10-6 Overview of Requirements for Care & Maintenance Tool Part Number Comments...
  • Page 582 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 10-6 System Maintenance 10-6-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary.
  • Page 583 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10-6-2 Functional Checks The functional checks take about 60 minutes to perform. Refer to the Voluson™ P8 / Voluson™ P6 Basic User Manual whenever necessary. 10-6-2-1 System Checks Table 10-8...
  • Page 584 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10-6-2-2 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to Table 3-2, ?Approved Peripherals,? on page 3-68 for a list of approved peripherals.
  • Page 585 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10-6-5 Physical Inspection Table 10-12 Physical Checks Step Item Description Labeling Verify that all system labeling is present and in readable condition. Scratches &...
  • Page 586 The Basic User Manual and/or care card provides a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. Review the Basic User Manual of Voluson™...
  • Page 587 To guarantee that the sensitive probe head does not hit the bottom of Endocavity soaking cup kit RIC-Holder is available at GE Accessory the container, place the probe in the RIC Holder during the disinfection (Order Nr. : E8385TA ) procedure.
  • Page 588 In case of unexplainable abrupt changes of safety-test results consult experienced !! CAUTION: authorized service personnel or GE for further analysis. To minimize the risk of a probe causing electrical shock, the customer should observe the following recommendations: •...
  • Page 589 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10-8-2 GEHC Leakage Current Limits for Voluson™ P8 / Voluson™ P6 The following limits are summarized for IEC60601-1 standard including national deviations and IEC 62353 Medical Electrical Equipment - Recurrent test and test after repair of medical electrical equipment.
  • Page 590 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Country Conditions Others Close Protective earth and close neutral with reverse polarity 0.01 mA* 0.01 mA*** Open Protective earth and close neutral with normal polarity 0.05 mA* 0.05 mA*** Open Protective earth and close neutral with reverse polarity...
  • Page 591 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 10-18 Type CF Patient Auxiliary Limits Country Conditions Others Close Protective earth and close neutral with normal polarity 0.01 mA* 0.01 mA*** Close Protective earth and close neutral with reverse polarity 0.01 mA*...
  • Page 592 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL NOTE: * Measurement limits per UL 60601-1 Medical Electrical Equipment Safety Standards Table IV ** Measurement limits per UL 60601-1 Medical Electrical Equipment Safety Standards, Table 19.5DV.1 *** Measurement limits per IEC 60601-1 Ed 2.
  • Page 593 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10-8-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
  • Page 594 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10-8-4 Grounding Continuity Electric Shock Hazard! CAUTION The patient or operator MUST NOT come into contact with the equipment during this test !! CAUTION: Measure the resistance from the third pin of the attachment plug to the exposed protectively-earthed metal parts of the case.
  • Page 595 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10-8-5 Enclosure Leakage Current Test 10-8-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible non- protectively-earthed parts of equipment if the ground wire were to break.
  • Page 596 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL ii)Others Hot : Brown Neutral : Blue Ground : Green-yellow Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer tool compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
  • Page 597 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10-8-5-3 Data Sheet for Enclosure Leakage Current The test passes when all readings measure less than the value shown in Table 10-14.
  • Page 598 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10-8-6-2 Generic Procedure The testing meter is connected from accessible protectively-earthed parts of the equipment to ground. Measurements should be made with power line polarity normal and reversed, and with the neutral open and closed.
  • Page 599 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 10-24 Typical Data Sheet for Earth Leakage Current Measured Earth Leakage Tester Neutral Tester Polarity Switch Current Enter Name of tested peripheral here Closed Normal Closed...
  • Page 600 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Neutral : White Ground : Green or Green-yellow ii) Others Hot : Brown Neutral : Blue Ground : Green-yellow Table 10-25 Typical Data Sheet for Type CF Patient Leakage Current Test - ECG Leads Measured Earth Leakage Tester Neutral...
  • Page 601 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10-8-8-2 Generic Procedure on Leakage Current Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with neutral open and closed.
  • Page 602 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Table 10-27 Typical Data Sheet for Type BF Patient Auxiliary Current Test Measured Leakage Tester Neutral Tester Polarity Switch Ground Switch Current Closed Normal...
  • Page 603 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Measured Leakage Tester Neutral Tester Polarity Switch Ground Switch Current Closed Reversed Closed 10-8-12 Total Patient leakage current : Type CF probe to ground(Only for Europe) 10-8-12-1 Definition This test measures the current that would flow to ground from all ECG leads through a patient who is...
  • Page 604 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL 10-8-13 Total Patient leakage current : Mains to Type CF probe (Only for Europe) Reference the procedure in the IEC 60601-1 Ed 3. Measure leakage current flow from mains to all ECG leads connected together.
  • Page 605 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Section 10-9 When There's Too Much Leakage Current... 10-9-1 Earth and/or Enclosure Fails Follow the below checklist in case of earth and/or enclosure leakage current failure. •...
  • Page 606 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL ULTRASOUND INSPECTION CERTIFICATE Customer Name: System ID: Dispatch Number / Date Performed: Warranty/Contract/HBS System Type Model Number: Serial Number: Manufacture Date: Probe 1: Frequency: Scan Format*:...
  • Page 607 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL ELECTRICAL SAFETY Outlet and Ground Continuity test Max Value Value Outlet/Ground Continuity Comments Allowed Measured Outlet (correct ground &wiring config.) System Ground Continuity 0.2 Ohm Earth leakage current test...
  • Page 608 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Type CF Patient Leakage Current Test - ECG Leads ECG lead 1 Tester Tester Polarity Max Value Value Ground Switch Comments Neutral Switch...
  • Page 609 DRAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Type BF Patient Leakage Current Test - Probes (This test should be done for all probe) Tester Tester Polarity Max Value Value Ground Switch Comments...
  • Page 610 RAFT ™ P8 / V ™ P6 OLUSON OLUSON DIRECTION 5723243, R DRAFT (M 23, 2018) EVISION ASIC ERVICE ANUAL Total Patient Leakage Current - Type CF probe to Ground Tester Tester Polarity Max Value Value Ground Switch Comments Neutral Switch Allowed Measured...
  • Page 612 Voluson™P8 / Voluso™P6 © 9, Sunhwan-ro 214beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea www.gehealthcare.com...

This manual is also suitable for:

Voluson p6 bt165670023567002456699995670021

Table of Contents