™
Carestation
620/650/650c (A1)
Standby
3-12
When in Standby, most of the alarms are disabled. The Standby
screen shows after End Case is selected.
If the fresh gas flow is more than 300 ml/min, a medium priority
alarm 'Turn off gas flow' occurs reminding the user to turn off the
fresh gas flow.
From the Standby menu:
•
A case can be started using the default settings by increasing
the fresh gas flow, setting Bag/Vent switch to vent, turning on
Aux O2+Air switch, or by turning on ACGO switch.
•
Select System Setup to open the System Setup menu.
•
Select Checkout to open the Checkout menu.
•
Select any other active function key to open the Start Case
menu.
2076152-001