Circuit O2 cell installation

2076152-001
Use the following instructions to add an O2 cell to the system.
Note
If an O2 cell is already in the system and needs replacement, see
the "Circuit O2 cell replacement" section.
1.
Install the O-ring onto the adapter and screw the adapter onto
the O2 cell.
1
1.
O2 cell
2.
Adapter
3.
O-ring
2.
Pull down the canister lifter handle to unlock the canister from
the breathing system.
3.
Lift the canister and remove it from the canister lifter base.
4.
Open the breathing system door. Push the release button to
unlock the breathing system.
5.
Lower the breathing system.
8 Setup and connections
2
3
8-11