620/650/650c (A1)

Airway module specifications

Gas specifications for airway modules

Airway humidity (patient
spirometry)
Sampling delay
Total system response time
Warm-up time
Respiration rate
Diversion flow
Airway pressure
Accuracy during stable conditions
Ambient temperature: 10 to 40°C
Ambient pressure: 495 to 795 mmHg
Ambient humidity: 10 to 98% RH, non-condensing
Automatic compensation for ambient pressure
Full module accuracy for Respiration rate of 4 to 70 breaths/min
CO2
O2
N2O
11-20
Use only airway modules that have anesthetic agent monitoring and
O2 monitoring on this system. The following modules can be used on
this system:
Airway Gas Option: N-CAiO
CARESCAPE
.
10% RH to 100% RH
3.0 seconds typical with a 3 m sampling line
Less than 3.8 seconds with a 3 m sampling line
1 minute for operation with CO2, O2, and N2O
5 minutes for operation of anesthetic agents
20 minutes for full specifications
4 to 100 breaths/min
120 +/- 20 ml/min
-20 cmH2O to 100 cmH2O
.
+/- (0.2 vol% + 2% of reading)
+/- (1 vol% + 2% of reading)
+/- (2 vol% + 2% of reading) between 0 and 85 vol%
+/- (2 vol% + 8% of reading for N2O between 85 and 100 vol%
series: E-sCAiO and E-sCAiOV