Maintenance Summary And Schedule - GE Carestation 620 User's Reference Manual

Table of Contents

Advertisement

Carestation
620/650/650c (A1)

Maintenance summary and schedule

CAUTION
Minimum Frequency
Daily
Weekly
Monthly
During cleaning and setup
9-4
These schedules indicate the minimum frequency of maintenance
based on typical usage of 2000 hours per year. Service the
equipment more frequently if it is used more than the typical yearly
usage.
Please contact an authorized representative of the manufacturer for
information concerning the decommissioning of equipment.
Equipment and accessories must be disposed of in accordance with
applicable regulatory and hospital requirements in effect at the time
and place of disposal.
Note
Local policies or regulations may require that maintenance be
performed more frequently than stated here.
Note
An additional level of disinfection is available. Contact an authorized
service representative to replace bacterial filters and to remove the
ventilator waste gas passage (gas to the system AGSS) for
disinfection.
Do not wipe the surfaces or clean the system using
organic, halogenated, or petroleum-based solvents,
anesthetic agents, glass cleaners, acetone, or other
harsh cleaning agents. Doing so can damage the labels
or the system. Refer to the "Cleaning and Sterilization"
manual for additional cleaning information.
.
Maintenance
Clean the external surfaces.
Replace the absorbent in the canister (as needed).
Perform weekly O2 cell calibration.
Perform monthly O2 cell calibration.
Perform a flow and pressure calibration.
Inspect the parts for damage. Replace or repair as necessary.
2076152-001