Carestation
620/650/650c (A1)

Moving and transporting the system

WARNING
8-4
The system may overbalance or tip if tilted more than 10
degrees.
Do not exceed specified load ratings when mounting
equipment on the top of the machine or on any of the
side rails.
See "Attaching equipment to the top of the machine" and
"Specifications and theory of operation" for system load ratings.
Before moving or transporting the anesthesia machine, place the
display arm and shelf in the transport position. Make sure that the
arm and shelf remain in position while moving the system.
The system may tip if the display arms are not in the transport
position while moving the machine on an incline. Remove the patient
monitor or hold the display in the transport position while moving the
machine.
Before removing the system from ceiling mount or wall mount,
remove any extra load attached to the system.
When using the brake on an incline, be sure to lock the brake with
the machine facing down the slope.
2076152-001