System Settings

After entering Setup Mode, press YES to scroll to System Settings, and then press RETEST to enter
the System Settings menu.
Volume
Options:
Lowest, Low, Med, High, And Highest
Default:
Med
Access Level:
General
The volume setting controls both the key beep and the voice prompt volumes. To change the volume,
follow the steps below.
1.
Press RETEST to enter the Volume option. The current setting blinks.
2.
Press either YES or NO to scroll to a new setting, and press RETEST to accept the setting. A
sample sound clip plays. If the volume level is not satisfactory, press RETEST to reenter Volume
and change the setting.
You can scroll to the next System Settings option -or- press MANUAL to return to the Main
menu, or press START + YES to exit Setup Mode.
Scale Prompt
Options:
OFF, every time for or any combination of days ( _ _ Mo _ _ We _ _ Fr _ _), or only one
time each day ( _ _ Mo* _ _ We* _ _ Fr* _ _ )
Default:
Every day ( Su Mo Tu We Th Fr Sa)
Access Level:
General
Scale Prompt allows you to turn the audio and text scale prompts off or turn the prompts on for all
or any combination of days. The factory default is on for every day. If the prompt is turned on, the
monitor prompts for (and collects) the patient's weight at the beginning of each monitoring session.
In the System Settings menu, scroll to Scale Prompt, and press RETEST to enter. The first
selection blinks. If the prompt is off, only underscores display.
1.
Press either YES or NO to change the setting, and press RETEST to accept the new setting.
2.
Continue through the seven settings until all are set correctly.
P.4510EN.01
GENESIS DM CLINICIAN GUIDE
Programming the Genesis DM Monitor
12/5/2007
24