Access Level:
General
You can use this menu to set the initial NIBP cuff inflation setting on a single-user monitor. (Otherwise,
set the initial NIBP on the patient ID card.)
Genesis DM uses this initial setting only for the first blood pressure collection because the monitor
stores the latest BP reading in memory to use as a setting for the next blood pressure. The monitor
inflates the cuff to 35 mmHg higher than the last systolic reading.
Once the monitor is powered down the default setting of 175 mmHg is re-established.
In the System Settings menu, scroll to Inflation Settings, and press RETEST to enter. The current
setting blinks.
Press either YES or NO to scroll to the desired inflation setting, and press RETEST to accept the
new setting.
You can scroll to the next System Settings option -or- press MANUAL to return to the Main
menu, or press START + YES to exit Setup Mode.
Display Brightness
Options:
Low, Med, High
Default:
High
Access Level:
General
This menu allows you to adjust the brightness level of the VFD display. To adjust the brightness level,
follow the steps below.
1.
In System Settings, scroll to Display Brightness, and press RETEST to enter. The current
selection blinks.
2.
Press either YES or NO to scroll to the desired brightness, and press RETEST to accept the new
setting.
You can scroll to the next System Settings option -or- press MANUAL to return to the Main
menu, or press START + YES to exit Setup Mode.

Questions Menu

The Questions Menu allows you to program the monitor to ask subjective disease-state questions
similar to the questions you might ask during a patient visit. There is a block of fifty-one standard
P.4510EN.01
GENESIS DM CLINICIAN GUIDE
Programming the Genesis DM Monitor
12/5/2007
26