Honeywell GENESIS DM Clinicians Manual page 62

Hide thumbs Also See for GENESIS DM:
Table of Contents

Advertisement

Patient training
The Genesis DM begins the process of uploading data as soon as it detects the device. The audio and
text prompts inform the patient that the upload is in progress and then announces when the data upload
is finished. If the patient fails to attach the device and the monitor times out, Genesis DM moves on to
onto the next scheduled activity in the monitoring session. The following table contains glucose meter
prompts.
Condition
Genesis DM is ready to initialize communication
If this is the first time this Genesis DM has received
information from this glucose meter
Transmission
The upload is complete
The Genesis DM monitor does not display glucose test results.
Glucose meter specifics
Some glucose meters require that the patient manually turn them on before they are able to transmit
data. The text message that displays on the face of the glucose meter differs from meter to meter. The
following table explains important differences between glucose monitors.
Meter
LifeScan One Touch Basic
LifeScan One Touch Profile
LifeScan One Touch Ultra
Bayer Ascensia Elite XL
Roche Accu-Chek
Home Diagnostics Prestige
The following table contains important notes specific to glucose meter types.
Meter
Bayer Ascensia Elite XL
LifeScan One Touch
Roche Accu-Chek
P.4510EN.01
GENESIS DM CLINICAL GUIDE
Text and audio
Please attach the glucose meter to the monitor.
Please make sure the cable is attached to the port marked COM (1, 2 or A) on the back of the
monitor.
New glucose meter detected
Receiving Data, please wait
Data upload is finished
Turn on to
Glucose meter text display during transmission
transmit?
Yes
Code ( ) Insert strip
No
PC
Yes
- - - MEM followed by a count down of test results
No
No display (unit does not turn on)
Normal display screen shows during transmission, device turns off when transmission
No
Note
The meter must be OFF when you attach it to its connector cable. If
not, the device will not transmit readings.
Until the patient turns the meter on, the Genesis DM monitor
CANNOT detect the device. The monitor will prompt the patient to
verify the cable connection to the com port. However, as soon as the
device is on and monitor detects it, the prompt will stop.
The meter does not transmit readings when the batteries are low.
Instruct the patient to follow the manufacture's recommendations and
instructions to maintain battery power.
The meter must remain off during transmission. It is also important
that the patient does not press any buttons during transmission.
12/5/2007
Patient Training
complete
58