Time Menu

Enter Setup Mode and, if necessary, scroll until Time Menu displays on the monitor. Press RETEST
to enter the Time menu. Use YES or NO to scroll to the Time Menu option you wish to enter.
Report Time (1 – 4)
Options:
Time (in selected format), and Off
Default:
Time 1 defaults to 8:00; all other times are set to Off
Access Level:
General
The Report Time is the scheduled time for the Genesis DM monitor to collect and transmit patient
vitals. Patients can have up to four report times in one 24-hour period. If you set more than one time,
make sure that you schedule report times at least one hour apart to accommodate the Scheduled
Reading window in LifeStream. For more information about scheduled readings, refer to the LifeStream
Clinician Guide.)
The Report Time displays in the format programmed in the Time Format menu (a.m. /p.m. or 24-hour
format).
If you assign multi-use status to a monitor, make sure all of the monitor's Report Times are set to Off.
In the Time Menu, scroll to the Report Time option that you wish to set (1, 2, 3, or 4).
1.
Press RETEST to enter. Depending on the current setting, either On or Off blinks.
2.
To change the setting, press either YES or NO.
3.
Press RETEST to accept the setting. The hour blinks.
4.
Scroll to the correct hour. Press RETEST to accept the hour setting. The minutes blink.
5.
Scroll to the correct minutes. Press RETEST to accept the minute setting.
You can scroll to the next Time Menu option, press MANUAL to return to the Main menu, or
press START + YES to exit Setup Mode.
Report Time # (1- 4) Days
Options:
Every day or any combination of days
Default:
Every day ( Su Mo Tu We Th Fr Sa)
Access Level:
General
P.4510EN.01
GENESIS DM CLINICIAN GUIDE
Programming the Genesis DM Monitor
12/5/2007
20