Temporal Temperature - Honeywell GENESIS DM Clinicians Manual

Hide thumbs Also See for GENESIS DM:
Table of Contents

Advertisement

Temporal Temperature

The Exergen TemporalScanner, available only from Honeywell HomMed, can transmit temperature
readings to the Genesis DM monitor.
The scanner is activated when the patient presses the red ON button and continues to scan and
transmit until the patient releases the button. The device scans for and transmits the highest
temperature so the last reading is always the highest temperature reading.
The monitor waits five seconds after the receipt of the last temperature reading. If no further readings
are received within the five second period, the last reading is bundled into the data packet for
transmission to the LifeStream.
If the patient resumes using the scanner within the five-second period, the monitor waits until readings
once again stop, allows another five-second grace period, and then bundles the last reading received into
the data-packet.
Assigning a COM port
You must assign a COM port to the scanner or the monitor will not prompt the patient for data or
communicate with the device to collect data (even if the device is connected to the COM port).
Assigning a COM port triggers the monitor to include temperature prompts during the monitoring
session and to communicate with the scanner.
The following instructions assume you know how to navigate in Setup Mode. If you have not
programmed a monitor, refer to Section 3 Programming the Genesis DM Monitor for complete
programming instructions.
The TemporalScanner requires an isolated port (COM-1 or COM-2). To assign the COM port, follow
the steps below.
1.
Enter Setup Mode, scroll to and enter the Option Menu.
2.
Scroll to the External Temp menu and press MANUAL.
3.
Assign COM-1 or COM-2.
4.
In the External Temp Time menu, set the collection frequency.
5.
Exit Setup Mode.
Attaching the TemporalScanner to the monitor
The scanner has a cord with a 4-pin mini-DIN connector that plugs into a Genesis DM com port.
To attach the scanner:
P.4510EN.01
GENESIS DM CLINICAL GUIDE
12/5/2007
Patient Training
67