Calibration:
Factory calibrated over the range of 70% to 99% SpO2 using human blood samples to functional saturation. Test
methods available upon request. No in-service calibration is required.
Sensor:
Red - 660nm, 2mW (typical)
Infrared 905nm, 2-2.4mW (typical)
Display Update Rate:
1Hz (Maximum age of SpO
Heart rate (oximeter)
Range:
30-254 BPM (1 BPM increments)
Accuracy:
±2% or 2 BPM, which ever is greater
Alarms:
None. Not for continuous monitoring
Averaging:
8 second average
Display Update Rate:
1 Hz (Maximum age of SpO
Heart rate (NIBP)
Range:
40 – 200 BPM (adult mode)
Accuracy:
+ 5 BPM or 5%, whichever is greater
Pulse Strength (requires oximeter)
Range:
0-6 segment bar graph, indicates logarithmic strength of patient's pulse
Display Update Rate:
60Hz
Scale
Manufacturer:
Honeywell HomMed
Range:
0 to 500 lbs. (0 to 227 kg)
Accuracy:
+ 0.5 lbs. (+ 0.2 kg) or + 0.5%, whichever is greater
Update Rate:
< 0.5 seconds
Stability of Calibration:
Zero calibration is performed at power-up. Multipoint span factory calibration is preserved through power cycles.
Factory calibration is good for 5 years of normal use

Appendix F: Electromagnetic Guidances

Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The Genesis DM is intended for use in the electromagnetic environment specified below. The customer
or the user of the Genesis DM should assure that it is used in such an environment.
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/ Flicker Emissions
IEC 61000-3-3
P.4510EN.01
GENESIS DM CLINICAL GUIDE
data is 20 sec.)
2
data is 20 sec.)
2
Compliance
Electromagnetic Environment - Guidance
The Genesis DM uses RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Class B
The Genesis DM is suitable for use in all establishments, including
domestic establishments and those directly connected to the
Class B
public low-voltage power supply network that supplies building
used for domestic purposes.
Complies