Honeywell GENESIS DM Clinicians Manual page 87

Hide thumbs Also See for GENESIS DM:
Table of Contents

Advertisement

Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the Genesis DM
The Genesis DM is intended for use in the electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Genesis DM can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Genesis DM as recommended below, according to
the maximum output power of the communications equipment.
Rated Maximum Output Power
of Transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
using the equation applicable to the frequency of the transmitter, where
according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
P.4510EN.01
GENESIS DM CLINICAL GUIDE
Separation Distance According to Frequency of Transmitter
m
150 kHz to 80 MHz
d
P
= 1.17√
0.12
0.37
1.17
3.70
11.70
P
is the maximum output power rating of the transmitter in watts (W)
12/5/2007
80 MHz to 800 MHz
d
P
= 1.17√
0.12
0.37
1.17
3.70
11.70
d
in meters (m) can be estimated
Index
800 MHz to 2.5 GHz
= 2.33√
d
P
0.23
0.74
2.33
7.37
13.30
83