Cleaning
Clean the Primair 700 using a mild cleaning solution [e.g. dish washing liquid soap or the EPI-US product (P/N 1779065)] and a lint-free cloth or sponge agent. Rinse the surfaces thoroughly with clean water
(do not use any solvents). During cleaning, ensure that no water is allowed to enter the air pump unit. If liquid gets inside the breathing hose, make sure it is thoroughly cleaned and dried before using it with the
RPD. Take extra care not to get cleaning solutions or water near the air-inlet, the battery connector contacts or the gasket. The apparatus must not be immersed in fluid.
Dry all components and the exterior of the breathing hose with a clean, lint-free cloth and/or leave to air dry in a clean environment free of dust and particulates. Make sure all components are completely dry before
putting the breathing device back into service.
MAINTENANCE __________________________________________________________________________________________________________________________________________________________________________________________
Maintenance, cleaning and replacement or repairs shall be done only by personnel familiar with these user instructions and power assisted filtering RPDs, using parts supplied by the original equipment manufacturer.
Prior to use:
Functional inspection of the system
After use:
Cleaning and disinfecting the complete power assisted filtering RPD.
Every 6 years:
Replace the breathing hose o-ring.
STORAGE_________________________________________________________________________________________________________________________________________________________________________________________________
Inspect all components prior to storage. Store the RPD in a clean and dry environment as its original packaging or in a sealed container so that it is protected from sunlight, dust or any contaminants. The recommended
storage temperature for Primair 700 Series is -20°C – +35°C and the relative humidity is less than 90%.
TECHNICAL SPECIFICATIONS ____________________________________________________________________________________________________________________________________________________________________________
Operating conditions:
Temperature:
-12°C – +48°C
Relative humidity:
20-90%
Storage conditions:
Temperature:
-20°C – +35°C
Relative humidity:
< 90%
Blower unit:
Ingress Protection:
IP65 with battery, breathing hose, filter and filter cover installed
Housing/material:
PC/ABS
Motor life:
>10 000 hours
Noise level:
<75dBA
Flow rate (tight-fitting config.):
145 lpm
Flow rate (loose-fitting config.):
180-195-210 lpm
Hose connection:
bayonet
Weight:
805 g
Altitude:
0 – 3048 m above sea level
Explosive Atmosphere:
Do not use.
Connected Blower unit PA710CEU:
Radio frequency:
2.4 GHz
Battery:
Type:
Li-Ion rechargeable
Material:
Housing: PC/ABS
Latch:
Nylon
TROUBLE SHOOTING GUIDE _____________________________________________________________________________________________________________________________________________________________________________
PROBLEM
Blower will not turn on
Low air-flow alarm is active
Low air flow inside the respiratory interface
Low air flow inside the respiratory interface, but no alarm
Excessive air flow inside the respiratory interface, but no alarm
Caution symbol on display is constantly illuminated (confirmed
no low battery or low flow alarm)
Battery does not hold a charge
Blower unit is not able to pair with the SafetySuite (PA710CEU
only)
Gasket:
Weight:
Capacity:
Charging time:
Charging cycles:
Filter type
P R SL
A2 P R SL
* Based on laboratory testing.
Battery Charger:
Display:
Power Supply:
Single-bay Charger:
Belt and Comfort Pad / Back Pack Materials (Standard version):
Straps:
Back/Hip Support:
Comfort Pad:
Back pack:
POSSIBLE CAUSE
Battery is not sufficiently charged.
1) Filter has become saturated with contaminates.
2) Breathing hose is obstructed.
3) Cartridge air-inlets are obstructed.
Breathing hose is disconnected or blocked.
1) Breathing hose is partially blocked.
2) Respiratory interface is damaged.
Magnet is missing from the blower-side connector of the breathing
hose.
System fault.
1) Battery is nearing the end of its useful life or is damaged.
2) Battery charger is malfunctioning.
Blower unit does not emit the operational data wirelessly.
EN-4
Open-cell silicone foam
410 g
10.9V, 6700mAh, 73wH
4.5 hours
> 500 cycles
Minimum battery duration (hours)*
Loose-fitting interfaces
Default (low) flow
Medium flow
High flow
8
8
7-8
6.5 -7.5
6 -7.5
Real-time battery charge and health monitoring
AC 100-240V, 1.5A 50-60Hz
DC 19.0V, 2.37A
Woven Nylon
Thermoplastic Polyurethane, PC/ABS
Closed Cell Polyethylene Foam, Nylon Fabric (TPE Coated)
Thermoplastic Polyurethane (TPU)
CORRECTION
Replace battery with charged battery.
1) Replace the spoiled filter.
2) Check the breathing hose for blockages or other obstructions.
3) Check to ensure the filter air-inlets are clear of obstructions.
Check breathing hose to ensure there are no blockages inside.
Confirm both ends of the breathing hose are properly connected.
1) Check breathing hose to ensure there are no blockages inside.
2) Replace the respiratory interface.
Replace the breathing hose.
Return the defective unit.
1) Replace the battery with a new one.
2) Replace the battery charger.
Visit https://papr.honeywell.com/ for FAQ and other helpful
information, or contact our SafetySuite support team.