Plunger Wash Pump Logbook - Hitachi CHROMASTER 5160 Instruction Manual

Table of Contents

Advertisement

3.3.2 Plunger Wash Pump Logbook

The total running time after replacement of a plunger wash
pump (option) is indicated. This function is enabled only when
the plunger wash pump is connected. This function is used as
an index of a replacement for plunger wash pump.
Monitor screen
GLP
GLP
(LOGBOOK=1,
1
Select LOGBOOK
1
ENT
LOGBOOK
(PUMP SEAL=1,
2
A.VALVE=3,
Select P.WASH
2
ENT
LOGBOOK
(REPORT=1,
1
Select REPORT
1
ENT
Select RESET
2
ENT
P.WASH PUMP CHANGE
<2000 - 2100>
Reset logbook
NOTE: Enter the plunger wash pump replacing date.
Enter Month, Date, Year. And press ENT.
In other words, it can't be reset without pressing the ENT key.
Even if the previous data is still indicated, reenter a desired
date and press ENT . The total running time will be indicated
as zero, and integration will be started.
Monitor screen
PRESSURE M=2,
P.WASH=2,
D.MIX=4,
P.VALVE=5)
RESET=2)
P.WASH TIME
LAST CHANGED DATE
99999 h
The total running time after plunger wash pump
replacement displays "h".
ENT
or
ESC
Monitor screen
(MM DD YYYY)
9
1
2010
3 - 41
KEY LOCK=3)
09/01/2010