Hitachi CHROMASTER 5160 Instruction Manual page 658

Table of Contents

Advertisement

3.1 Basic Operation
3.1.4 Setting of Temperature Ready Conditions
Make the condition setting for the temperature ready status.
Initial setting: TOLERANCE=1.0 °C, WAIT TIME=0 min.
USE TOLERANCE (YES=1, NO=0)
0
TOLERANCE (°C)
<0.1 - 9.9>
TOLERANCE (°C)
<0 - 99>
NOTE: 1. Unless a tolerance range and wait time are entered
(and ENT key clicked) on the temperature ready
condition setting screen, the settings will not be
updated.
2. If "NO" is selected for "USE TOLERANCE", the
temperature ready status is provided regardless of
the current oven temperature.
3 - 8
COND
COND
Specify whether or not to
use tolerance setting.
1 ENT
WAIT TIME (min)
1.0
5
Set a tolerance range.
1.0 ENT
WAIT TIME (min)
1.0
5
Set a wait time.
5 ENT
Monitor screen

Advertisement

Table of Contents
loading

Table of Contents