3.5 Setups

3.5.2 Setting of Thermostatic Flow Cell

POWER
Press
and
The SETUP screen now appears.
SETUP:
DETECTOR
CH1
Press the [ENT] key.
The selection screen appears.
SETUP ITEMS (DETECTOR=1, THERMO CELL=2,
1
CONTRAST=3)
Press numerical key [2], and then the [ENT] key.
THERMO CELL (NO USE=0, USE=1)
0
For avoiding use of
thermostatic flow cell : Press [numerical] key [0], and then the
For selecting use of
thermostatic flow cell : Press [numerical] key [1], and then the
After setting use or non-use of the thermostatic flow cell, "ON" or
"OFF" is indicated on the monitor screen 2.
WL
SAMP
250
301000
When the thermostatic flow cell is not yet connected,
"---" will be indicated on the monitor screen 2.
WL
SAMP
250
301000
New setting is accepted after turning on power supply again.
NOTE: Set the thermostatic flow cell in accordance with
measuring conditions. (Recommended flow rate
range when using thermostatic flow cell: 2mL/min
or less).