Setting the Installation or Expiration Date for New Consumables

When you replace a consumable, you must enter the date of installation or expiration to ensure
that the Analyzer keeps an accurate record of usage and that the indicators on the Main screen
remain accurate.
1.
Press the Menu button and select the Maintenance tab.
2.
Press the Consumables button.
3.
Press the New... button that corresponds to the consumable you are installing.
• If you choose New Lamp, New Tube, or New RBed, press the Accept button to enter
today's date, or enter values for the appropriate Day, Month and Year and then press
the Accept button.
• If you choose New Acid, enter the expiration date that was listed on the acid
packaging: press the Day button and use the number pad to specify the day of
expiration and press Enter; repeat for the Month and Year buttons. Press Accept after
you have entered the complete expiration date. You will be prompted to confirm the
installation of new acid; press the Confirm button to continue.