Turbo Mode Maintenance Considerations

All regular maintenance functions described in Chapter 7, "Maintenance," apply to Analyzers
working in Turbo mode. In addition, if the Analyzer operates continuously in Turbo mode, the
operating life of the sample pump is reduced by half, to 2.5 years.
Table 19: Replacement Schedule for Consumables in Turbo Analyzers
Description
DI Water Reservoir
Oxidizer
Acid
UV Lamp
Sample Pump Tubing
Resin Bed
Sample Pump
GE Analytical Instruments ©2009
Typical Operating Life
Refill as needed (check when replacing reagents or UV lamp)
3 calendar months maximum (depends on flow rate)
12 calendar months maximum (depends on flow rate)
6 months
12 months
Typically 12 Months (depends on water quality)
2.5 years
224 of 236
DLM 90488-01 Rev. A