Make sure the DI water Reservoir is full, particularly when running samples with
high TOC or high salt concentrations. Always "clean-up" the Analyzer by running
low-TOC DI water after running high TOC or salt samples.
This is a Class A product. In a domestic environment, this product may cause
electromagnetic interference in which case the user may be required to take
adequate measures to correct the interference.
For performance within specifications on ozonated water systems, an Ozone
Destruct Kit must be purchased from GE Analytical Instruments and installed
according to instructions.
To avoid false TOC readings and possible damage to the Analyzer, always make
sure the sample inlet is open and the DI water reservoir is filled before starting
analysis.
GE Analytical Instruments ©2009
Warning
Warning
Warning
Warning
24 of 236
DLM 90488-01 Rev. A