GE Sievers 900 Series Operation And Maintenance Manual page 199

On-line total organic carbon analyzers
Table of Contents

Advertisement

21 - Tubing Warning
Message: "The tubing needs to be replaced."
Explanation: Based on the history of usage, the Analyzer estimates that the sample pump tubing
has expired. Order new pump tubing from GE now."
22 - Resin Bed Warning
Message: "The resin bed needs to be replaced."
Explanation: Based on the history of usage, the Analyzer estimates that the resin bed has expired.
Order a new resin bed from GE now."
24 - ICR Vacuum Decrease
Message: "Inspect data for instability. If ICR fault light remains lit for more than 30 minutes,
refer to the 900 ICR Manual."
Explanation: This error generally displays when the ICR is not creating a vacuum but is set to Inline.
Refer to the 900 ICR Operation and Maintenance Manual for troubleshooting suggestions.
25 - History Data Warning
Message: "A problem has been corrected in the history data. Contact GE if the warning persists."
Explanation: The Analyzer has detected a problem reading or writing to the data history memory;
as much of the data history as possible is reconstructed. If the problem continues, contact GE
Analytical Instruments technical support.
26 - System Restart Warning
Message: "The system has experienced an error and has automatically restarted. Contact GE if the
warning persists.
Explanation: The Analyzer has restarted for a reason other than a power failure. If the problem
continues, there may be a problem with the Analyzer and you should contact GE Analytical
Instruments technical support.
41 - System Restore Error
Message: "Failed to restore system."
Explanation: The attempt to restore the system settings failed. Make sure the USB drive that
contains the saved system settings is attached to the Analyzer properly and try to restore again. If
the error occurs again, try copying the files on the USB drive to a different USB drive, and then
attempt to restore system settings again using the new USB drive. If the error message persists,
contact GE Analytical Instruments technical support.
49 - USB Error
Message: "A problem has been detected with the USB device. Please refer to the 900 Oper. and
Maint. manual for info on supported devices."
Explanation: The Analyzer was unable to write to or read from the USB flash memory drive. Make
sure the USB device is attached to the Analyzer properly and attempt the operation again. If you are
not using the USB flash memory drive that was supplied in the Analyzer's accessories kit, it is
possible your USB device is not compatible with the Analyzer. For a list of known compatible USB
GE Analytical Instruments ©2009
199 of 236
DLM 90488-01 Rev. A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents