Exporting And Printing Constants; Set Up Data I/O - GE Sievers 900 Series Operation And Maintenance Manual

On-line total organic carbon analyzers
Table of Contents

Advertisement

1.
Select the I/O tab and press the Printer button.
2.
Press the Printer button and select your printer model, either Citizen or Seiko. If you do
not have a printer, make sure No Printer is selected.
3.
Press the Header Freq. button and select First Page to print a header only on the first
page of output or select All Pages to print a header on all pages of output.
4.
Press the Print Freq. button and select how often you would like to print TOC data
information.

Exporting and Printing Constants

Calibration constants, reagent flow rates, and other key parameters are stored in the Analyzer's
memory. You should export and print the factory settings for future reference.
To export the settings to a comma-separated text file, follow these steps:
1.
Select the Maintenance tab.
2.
Press the Advanced button.
3.
Press the USB I/O button. Make sure that the USB flash memory drive from the Analyzer's
accessory kit is attached to the USB port, then press the Save System button. Archive the
exported files in a secure location on your computer.
4.
Press the Back button and then press the Menu button.
If you have a printer attached to the Analyzer, you can print these settings for future reference by
following these steps:
1.
Select the Maintenance tab.
2.
Press the Advanced button.
3.
Press the Print Constants button.
4.
Press each of the four buttons, one at a time, to print the different constants.

Set Up Data I/O

If you installed binary input, alarms, or the 4-20 mA output, configure settings for those features
now. Select the I/O tab and configure the appropriate settings. Note that you may need to consult
GE Analytical Instruments ©2009
85 of 236
DLM 90488-01 Rev. A