Autoreagent calibration standards sets include the standards shown in Table 15.
Table 15: Standards Required for Autoreagent Calibration
Reagent Water
25 ppm TOC (as KHP)
50 ppm TOC (as nicotinamide)
25 ppm IC (as Na
*Note: 1ppm = 1 mg C/L, 1ppb = 1 µg C/L
Autoreagent verification standards sets include the standards shown in Table 16.
Table 16: Standards Required for Autoreagent Verification
Reagent Water
25 ppm TOC (as sucrose)
50 ppm TOC (as sucrose)
50 ppm IC (as Na
*Note: 1ppm = 1 mg C/L, 1ppb = 1 µg C/L

Preparing for Calibration

Export Current Constants

Prior to performing any calibration procedure, export the current constants, in the event they need
to be re-loaded or referred to in the future. To export the constants, follow these steps:
1.
Select the Maintenance tab.
2.
Press the Advanced button.