Printing Data History

The Analyzer can print the data history for any valid time range that you specify. To print data,
follow these steps:
1.
Select the Data tab.
2.
Press the Print button.
3.
Press the Range button to specify a preset time range: 1 Day, 1 Week, All, or Custom.
• If you select a Custom range, press the To Date and From Date buttons and enter the
desired time range values.
4.
Press the Values button to specify which data will print:
• TOC/IC/Mode — Prints TOC and IC data and the mode in which the data was measured.
• TOC/IC/TC — Prints TOC, IC, and TC data.
• TOC/Reagents — Displays TOC data and reagent flow rates.
• TOC — Displays TOC data only.
5.
Press the Print button to print the data.
Note:
Prior to printing, you may want to confirm your printer settings. See "Configuring
Printer Settings" on page 109 for more information.

Exporting Data History

The Analyzer can export the data history for any valid time range that you specify, either to the
serial (RS-232) port or the USB port.
To export data, follow these steps:
1.
Select the Data tab.
2.
Press the Export button.
3.
Press the Range button to specify a preset time range: 1 Day, 1 Week, All, or Custom.
• If you select a Custom range, press the To Date and From Date buttons and enter the
desired time range values.
4.
Press the Export button. Make sure a USB device or serial cable is attached to the Analyzer
prior to pressing the Export button.