1. After choosing Set on the previous screen, the new tag under Value is highlighted. You have
made a change on the Pocket PC, but it has not been downloaded to the device yet.
2. Click on the Upload button (short Arrow on previous screen) to send the new tag to the device.
3. You can choose the next button to the right of the Upload button (long arrow on previous screen)
if you want to clear the new tag value and revert back to the current value (STT25H) in the device.
The following screen shows an example of a Method. A method, identified by a blue icon, is a step-by-step
procedure to execute a change. For, example the SDC 625 uses a Method to change the Engineering units
or to put a pressure transmitter into output mode.
1. The method above allows one to change the engineering units of a field device.
When you have fully configured the device or made the changes you need to make, you can exit the
SDC625 program in the following way: Click on Device, and then Exit.