Procedural Considerations; Input Calibration - Honeywell MC ToolKit User Manual

Communications tool
Table of Contents

Advertisement

Procedural Considerations

Menu Item
Local Meter
(ST 3000 only)
Sensor
(STT 3000)
EXTENDED INFO
DYNAMIC VARS
BURST MODE
SPEC. MONITOR
Procedural Considerations
The details of procedures vary with device type. This section contains a set of procedures for Honeywell
HART Transmitters, and separate set of procedures for non-Honeywell HART devices.

Input Calibration

Input calibration of transmitters should be done only when necessary, and should be done only under
conditions that will ensure accuracy:
The transmitter should be taken out of service, and should be moved to an area with favorable
environmental conditions: clean, dry, and temperature-controlled.
The source for the input pressure must be very precise, and must be certified for correct operation.
The procedures should be done by qualified personnel.
For Honeywell HART devices, input calibration procedures are given in Table 12 through Table 14, and
for non-Honeywell HART devices, consult the manufacturer's User Manual..
Page 40
Select:
Meter Units (EU)
Enter:
Custom Units (conversion constant)
(Custom) EU Flow: Low/High
Select:
Sensor Type
Line Filter
CJ Mode Comp
Mode Selection (STT25T)
Match PV's (STT25T)
Enter:
Primary Variable Code
Secondary Variable Code
Tertiary Variable Code
Quaternary variable Code
Select:
Burst Mode
Burst Options
Select:
No. of Var's Query
Device variable
34-ST-25-20 MC Toolkit User Manual
Task
Observe (Read):
Meter Type
Observe (Read):
RTD Wire Selection (STT25H)
Observe (Read):
Universal Rev.
Software Rev.
Field Device Rev.
Poll Address
PROM ID
# Req. Preams
PV Sensor S/N
Final Assembly #
Observe (Read)_:
Values of selected variables
Release 3
9/06

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents