Procedural Considerations
Honeywell HART Diagnostics/Service Menu Procedures
Device
The Critical status block lists
Status
the conditions that render the
transmitter device inoperable.
The Non-Critical status
block lists advisories of
conditions that are
noteworthy, but that do not
cause non-operability.
Refer to the section on
Messages and Diagnostic
Codes.
Write
The
Write Protect
Protect
is available for the
Honeywell STT 25H
Temperature Transmitter.
Provides write protection for
all parameters.
To enable write protection,
select
Write protected
select the
Send
To disable Write Protection,
select Not write protected,
enter the password, and then
select the Send button.
Page 50
function
, then
button.
The password can be
changed only when Not
Write Protected.
To
Change Password
select the Start button, type
in the new password, and
then select the
34-ST-25-20 MC Toolkit User Manual
,
button.
Send
Release 3
9/06