The Circuit O2 Cell test measures the O2 percent.
To run this test, follow the on-screen instructions. Do not select
Done when 21 is first displayed. Allow the reading to stabilize, then
select Done. Calibrate the O2 cell if necessary.
When External Gas Monitor is set to Yes by the Super User, the
External Gas Monitor reminder occurs. This is not a test. This is a
reminder to connect a respiratory gas monitor.