Honeywell Apex Technical Handbook page 60

Gas detection system
Hide thumbs Also See for Apex:
Table of Contents

Advertisement

MAN0604_Issue 09_02-2013
2110M8030
4-20mA Set Zero
4-20mA Set Zero
ok
4-20mA Set Zero screen
Set zero value -
3.5mA to 4.5mA
ok
Calibration Menu
Force 4-20mA
Caution:
Ensure that when forcing a signal on the 4-20mA output that the effects on the
network and controller are understood.
Force 4-20mA
ok
Force 4-20mA screen
Set value - 0mA
to 22mA
esc
Calibration Menu
4. OPERATION
Use up/down
buttons
Use up/down
buttons
60
Calibration Menu
This option allows the This option allows
the zero point within the 4-20mA signal
range to be adjusted between 3.5
and 4.5mA. The default setting is 4.0mA.
The offset applies to all 4-20mA output
states including Inhibit, Warning, Overrange
and Alarms.
Calibration Menu
This option allows the 4-20mA output to
be forced to a level selected between 0
and 22mA. The default is 1mA. The forced
setting times out after approximately
5 minutes and the output returns to
normal active operation.
Apex