Installing The Gas Sensing Cartridge - Honeywell Apex Technical Handbook

Gas detection system
Hide thumbs Also See for Apex:
Table of Contents

Advertisement

MAN0604_Issue 09_02-2013
2110M8030
Set the link on the Interconnect PCB for the required Controller Area network (CAN) setting.
See 3.2.4.
(4)
Refit the top to the Certified Junction Box base.
Cautions:
1.
Ensure that there is no moisture inside the Certified Junction Box before fitting the lid.
2.
Use only the captive bolts supplied, replacement with alternative bolts will invalidate
certification.
The top should be located using the locating pins on the Certified Junction Box base and then
lowered onto the base. Ensure that the lid retaining cable and/or wiring are not trapped and
the O-ring in the top is correctly located. Check that there is no discernible gap between the
top and the base. Tighten the captive M8 bolts to 5Nm (3.68 foot-pounds).

3.2.3 Installing the Gas Sensing Cartridge

(1)
Remove the cap from the Certified Sensor body.
Rotate the cap or accessory 1/4 turn in an anticlockwise direction to release the
bayonet fitting.
(2)
Fit the gas sensing cartridge into the sensor body.
Each cartridge is provided with a certificate of calibration (printed on the reverse of the
instruction sheet, Part No: 2110M8015, supplied with the cartridge) that guarantees that the
cartridge is calibrated and ready for use. Before installing a cartridge check that the number
on the cartridge label matches the gas type and range for the function required.
3. inSTAllATiOn
Sensor Cartridges may contain corrosive solutions.
Dispose of according to local and national regulations.
Sensor
body
Sensor
cap
WARning