Procedure
1. On the acquisition PC's Persistence page,
click the Tools button.
The Calibration & Configuration menu ap-
pears.
2. In the Calibration & Configuration menu,
click Service.
3. When prompted, enter the service pass-
word (currently DDD) and click OK.
The Service Calibrations menu appears.
4. Click the High Voltage button.
Note. If the CardioMD AC option is installed on the system, all buttons of the Service menu, ex-
cept the Node configuration button, are greyed out. The transmission scanners must be removed
and the system configured to run without the AC option before any service calibrations can be
performed. See Chapter 12 CardioMD AC Option for instructions on removing the transmission
scanners.
1MAN0158-H11 04-2005 CSIP Level 1
Details
Calibration
CardioMD Service Manual 4-9