Philips CARDIOMD Service Manual page 441

Table of Contents

Advertisement

Procedure
9. Remove the CANbus terminator from
connector A3 on the EDC board on
detector 1.
10. Connect the AC Option's CANbus
interface cable (9CBL1314) to connector
A3 on the EDC board of detector 1.
11. On each detector, mount one of the female
DB15 connectors that are part of the AC
Option's detector interface cable
(9CBL1313) on the brackets provided.
Use the threaded stand-offs provided.
Note. The two female DB15 connectors that
are part of the AC Option's detector interface
cable carry identical signals. Hence, they are
not detector-specific.
1MAN0158-H11 04-2005 CSIP Level 1
Details
Threaded stand-off
CardioMD AC Option
CardioMD Service Manual 12-93