Installation
Procedure
1. Place a lock washer over each Allen
screw, concave side of washer facing the
screw head.
Caution. Make sure to place one lock washer
over each Allen screw. Leaving out the washer
entails a risk of damaging the detector.
2. Mount the 3 guide pins on the end piece of
each detector, taking care to mount the
coded guide pin towards the top of the
detector (where the EDC board is located).
The photo on the right shows detector 1 with
covers mounted.
Note. One of the guide pins for each detector
is coded and must be mounted at the top of the
detector, with its flat side facing away from
the table console. By doing this, you ensure
that the correct transmission scanner is placed
on each detector:
Transmission scanner 1 must be mounted
on detector 1 (facing detector 2), and
Transmission scanner 2 must be mounted
on detector 2 (facing detector 1).
12-102 CardioMD Service Manual
Details
Coded guide pin
This side away from
table console
1MAN0158-H11 04-2005 CSIP Level 1