Procedure

12.5.2.5 Replacing the Motor and Gear Assembly

Procedure
The photo on the right shows the motor and
gear assembly 9ACO1429. See also section
12.7.1.2, page 12-110.
1. Follow the instructions given in the
previous sections (section 12.5.2.3 and
12.5.2.4) to remove the motor power
board and the limit switches from the
motor and gear assembly.
2. Mount the motor power board and limit
switches on the new motor and gear
assembly.
1MAN0158-H11 04-2005 CSIP Level 1