Repair Procedures

12.5.3 Line Source Compartment Repairs

12.5.3.1 Removing Line Source Compartment Covers

To access the parts inside the line source compartment, both covers must be removed. The
following sections explain how to remove covers:
While the transmission scanner is mounted on the detector, and
When the transmission scanner is dismounted and the line source compartment has been
detached from the AC option bracket.
Transmission Scanner Mounted on Detector
Procedure
1. Use the hand controller to position the
detector assembly conveniently for access
to the transmission scanner you are going
to work with.
2. If required, use the crank supplied with the
CardioMD system to position the line
source compartment conveniently.
CAUTION
12-38 CardioMD Service Manual
Details
Caution. Make sure that all objects are clear of the detectors and transmission
scanners before activating hand controlled motions.
1MAN0158-H11 04-2005 CSIP Level 1