Procedure
14. When the Persistence page reappears,
press [F3].
15. The CardioMD application displays the
study files stored in the folder you have
just set up.
16. Select either the patient (uppermost list) or
the study/studies to be imported and click
Transfer.
More detailed operating instructions can be
found in the CardioMD Operator's Manual,
Chapter 4 Operating Instructions. See the
sections Exporting a Patient and Exporting a
Study.
17. When the data transfer is complete, click
Done to return to Persistence page.
The study files have been imported into the
patient database.
When done with the import operation, it is
important to delete the folder node:
18. Proceed as explained in step 1 – 5 to
display the Nodes Configuration dialog.
19. Select the Folder entry and click the
Delete button.
20. When prompted to confirm the delete
operation, click OK.
21. Click Exit.
22. Close down and then re-start the
CardioMD application to disable the
service password.