Repair Procedures
Procedure
2. Use a 2.5 mm Allen key to remove the
two screws holding the attenuator.
3. Pull out the attenuator in the direction of
the arrows.
CAUTION
Procedure
4. Remount the two screws to lock the
attenuator in the open position.
The photos on the right show the attenuator in
the open position.
5. Reinstall the line source housing inside the
line source compartment. See instructions
in the section Installing a New Line
Source, page 12-71.
6. After reinstalling the line source housing,
grease the acme screw inside the line
source compartment. See section 12.8.1,
page 12-116 for instructions.
7. Go through the procedure described in
section 12.6.5, page 12-109 to ensure that
the CardioMD system is ready for use.
12-76 CardioMD Service Manual
Details
Caution. The attenuator consists of four separate copper strips. Be careful to
retract all four strips, aligning the holes of each strip exactly with the holes in
the line source housing.
Details
1MAN0158-H11 04-2005 CSIP Level 1