Installation
Procedure
The gantry-PC interface cable must be
protected by means of approximately 1.5 m of
rubber cable protector, routed and secured to
the floor as indicated to the right.
Procedure
The CardioMD system is supplied with 3 m
(10 ft.) of rubber cable protectors, used for
covering cables running from the gantry to the
wall (power and network cables) and the cable
between gantry and acquisition PC.
Cable protectors should be applied where
traffic across the cables is to be expected.
7. Cut cable protectors to desired lengths.
8. Open the membrane on the base of the
protector to slide in the cable.
8-46 CardioMD Service Manual
Details
Acquisition
PC cart
WARNING
Trip hazard! Failing to protect and secure the gantry-PC interface cable may
cause cables to be damaged and persons to trip over the cable.
Details
Detector
Patient Table
Rubber cable
protector
1MAN0158-H11 04-2005 CSIP Level 1