Acquisition PC Software Updates

7.3 Acquisition PC Software Updates

The CardioMD software CD-ROM includes two batch files that are used to update or completely
install the acquisition software.
7.3.1

Upgrading CardioMD Acquisition Software

7.3.1.1

Tools Required

CardioMD software installation CD, part number 9CSY0906.
Note. As a precaution, in case anything should go wrong, ensure that the previous version of the
software CD-ROM is available in order to be able to restore the original acquisition software ver-
sion.
7.3.1.2

Procedure

Before starting the acquisition PC software update, check the release note on the CD-ROM. You
find the release note in the directory
that is more recent than this manual.
Procedure
Before starting the acquisition software up-
grade, close down the CardioMD application:
1. If the CardioMD application is not dis-
playing the Persistence page, click the Exit
button to return to the Persistence page.
2. On the Persistence page, click the Close
button in the upper right corner of the
CardioMD application window.
3. Close all other programs running on the
acquisition PC.
4. Insert the CardioMD software CD in the
acquisition PC's CD-ROM drive.
5. Start the Windows Explorer and select the
CD-ROM drive.
6. Run the file
InstallCardioMDup-
by double-clicking on the
date.bat
name of the file.
The file is located in the root directory of the
CD-ROM.
7. Follow the instructions given.
During the execution of the
may several times see a message asking you
whether existing files should be overwritten:
8. Enter
to indicate that all files should be
A
7-18 CardioMD Service Manual
Release notes
Details
file, you
.bat
. The release note may provide information
1MAN0158-H11 04-2005 CSIP Level 1