Procedure
4. Lift up the end of the cover as shown and
slide the cover all the way out.
CAUTION

12.5.3.2 Dismounting Connecting Cable

The 9CBL1344 cable connecting the bracket and the line source compartment can be replaced
when the line source compartment has been dismounted from the bracket and the line source
compartment covers have been removed.
Procedure
Note. The 9CBL1344 cable between the
bracket and line source compartment is
attached to the line source compartment by a
set screw.
1. Remove the set screw through the small
hole in the line source compartment
indicated in the photos on the right.
1MAN0158-H11 04-2005 CSIP Level 1
Details
Caution. When remounting the service cover, it is very important to be
cautious to avoid damaging the collision sensors under the cover. See section
12.5.3.10, page 12-77.
Details
CardioMD AC Option
CardioMD Service Manual 12-43