Installation
Next, mount wiremolds and route the cable to the gantry. Proceed as follows:
Procedure
1. Route wiremolds nicely from the desk-
mounted laptop to the gantry.
Use the adhesive on the wiremolds to
attach them to the wall.
The wiremolds provided with the system
include angles and turns needed for most
installations.
In case you need additional parts, these can be
obtained locally.
Note. The adhesive on the wiremolds is only intended as an aid when installing. The adhesive is
not sufficiently strong to keep wiremolds in place for longer periods of time.
Procedure
2. With a 6 mm drill, drill holes through the
wiremold and into the wall.
Note. Take care to avoid any studs.
3. Insert wall plugs into the holes drilled.
4. Use the screws provided with the system
to secure wiremolds.
5. Route the cable to the gantry inside the
mounted wiremold.
8-62 CardioMD Service Manual
Details
Details
WARNING
Never route the gantry-PC interface cable together with any other cables. Due
to safety and EMC/EMI considerations, the cable connecting the acquisition
PC to the gantry must be routed in separate wiremolds.
1MAN0158-H11 04-2005 CSIP Level 1