Detector Repairs
Procedure
the LED aperture should be changed.
Note. To change the LED aperture, turn the
slot on the back of the PMT. The slot can be
accessed with a screwdriver through a hole in
the PMT strip board.
18. Remount the rest of the detector and run a
complete PMT update calibration.
19. Run a QC check to determine if further
calibrations are required. See the
CardioMD Operator's Manual Chapter 5
Maintenance for instructions.
6.3.6
Replacing the Crystal
Procedure
1. Disassemble the detector and remove all
24 PMT's. Follow instructions provided in
the previous sections.
2. Unscrew the mu-metal support plates
mounted on the two short edges of the
detector.
3. Lift out the mu-metal support plates.
4. Disconnect the internal HV cable from the
mu-metal assembly.
5. Loosen the two set screws in the support
brackets in each side.
To remove the brackets:
6. Unscrew the 3 screws going through the
detector casting from the outside and the 3
screws going through the detector casting
from the power supply compartment.
7. Disconnect the internal HV cable from the
crystal and lift up the crystal.
8. Remove the gasket – or remainders of the
gasket – from the crystal seating inside
detector casting.
6-10 CardioMD Service Manual
Details
Details
mu-metal plate
1MAN0158-H11 04-2005 CSIP Level 1