Philips CARDIOMD Service Manual page 27

Table of Contents

Advertisement

I, II
Handset
PC
EKG
E-Stop
Figure 3.2 Gantry schematic block diagram, CardioMD Series I and II systems
On Series I and II CardioMD systems, the side panel contains all interface connectors to
I, II
peripheral devices, including the acquisition PC and the hand controller. A 115 V AC power
supply output is available for the acquisition PC, along with IEEE 1394 FireWire communication,
which is routed directly to the detectors. An input connector for an ECG trigger is also provided.
This R-wave input is also routed directly to the detector EDC modules where the trigger events
are encoded into the data stream going to the PC.
Interconnections of Series III and earlier systems are shown in the diagrams Figure 3.3 and Figure
3.4.
1MAN0158-H11 04-2005 CSIP Level 1
E-Stop
Safety Board
Handset
E-Stop
inputs
LED
Side Panel
115 V
AC
Ultimitate motion limit switches
(X, Y, Rotate)
EKG
Firewire
Collision
24 V
Handset
AC PSU
55 V
180 V
55 V
24 V
Motor
Motor
Motor
Stop
Detailed Description
To Detector
CANbus
Y Drive
I/O
180 V
Control
CHA
Amplifier
Position
Control
CHB
Amplifier
Position
55 V
X Drive
HSSI
CANbus
Position
Position
CHB
Position
ENC
180 V
Control
I/O
CHA
Amplifier
Position
In
I/O
Control
Out
24 V
I/O
Table Drive
CardioMD Service Manual 3-3
Absolute
Position
Encoder
Y-Position
Motor & Gear
X-Position
Motor & Gear
Absolute
Position
Encoder
Rotate Drive
Absolute
Position
Encoder
Rotate Position
Motor & Gear
Table
Motor
Mid-position
switch