Wiring The 4-20 Ma Connection; Table 3 Serial And 4-20 Ma Outputs (Tb3) - GE Sievers 500 RL Operation And Maintenance Manual

On-line total organic carbon analyzer
Hide thumbs Also See for Sievers 500 RL:
Table of Contents

Advertisement

Chapter 3: Installation
12
11
10
9
8
7
6
5
4
3
2
1

Wiring the 4-20 mA Connection

The data from the Analyzer may be recorded using the 4-20 mA outputs located on TB3. The analog output is
verified at the factory prior to shipping. Wiring connections are shown in Figure .
Internal to Analyzer
4-20 mA Output
Isolated self power
Voltage compliant to 20 V
GE Analytical Instruments ©2010
Table 3: Serial and 4-20 mA Outputs (TB3)
Pin Number
(-)
(+)
TB3
Figure 3: Wiring Diagram for the 4-20 mA Connection
Output
4-20 mA 1 (- Output)
4-20 mA 1 (+ Output)
4-20 mA 2 (- Output)
4-20 mA 2 (+ Output)
Reserved
Reserved
Reserved
4-20 mA 3 (- Output)
Serial (In)
Serial (Out)
Serial (ground)
4-20 mA 3 (+ Output)
Max. Load 600 ohms
No external power required
(-)
(+)
Recorder
51 of 226
or
Rload
DLM 74001-04 Rev. A