Chapter 6: Calibration and System Protocols
2.
If the Analyzer is taking measurements, stop analysis.
3.
Press the Menu button, select the Maintenance tab, press the Cal/Ver/Validate button, and then press
the JP Protocol button.
4.
If you have a Super IOS system, the Select Rinse screen appears. Select one of the following rinse
options, and then press the Next button:
• No Rinse
• Before
• After
• Before and After
5.
If you have selected No Rinse or After, and there are vials in the Super iOS, remove them now. Press
Next to continue. Wait for the Analyzer to drain each of the Super iOS vial chambers (about 2 minutes).
6.
If you have selected to run a rinse, the Analyzer will prompt you to do one of the following:
• Insert the Super iOS Rinse vial set into the Super iOS. Press Next to continue.
• Or, if using individual rinse vials, press Next for additional instructions. Insert the individual vials in
the Super iOS, and then press No Set to proceed.
7.
Do one of the following:
• If you have a purchased vial set, insert the JP Protocol standards cartridge into the Super iOS
System with the label facing away from the Analyzer and press Next. Proceed to Step 11.
• Insert the individual vials into the Super iOS system in the following order:
Port 1 = Leave empty
Port 2 = Reagent Water (Rw) Blank Standard
Port 3 = SDBS Standard
Port 4 = KCL Standard (Optional, only if the pharmacopeia is set to include JP TOC and Conductivity)
Press the Next button, and then press the No Set button. Proceed to step 9.
NOTE:
Open the cartridge or vials bag and avoid touching the septa to protect
against introducing foreign particles.
8.
If you have a standard iOS system:
• Slide the iOS system door open, and then wait for the water to drain (about 30 seconds).
• Open the vials bag and avoid touching the septa to protect against introducing foreign particles.
• Insert the Reagent Water (RW) standard into the iOS System, and then press Next.
• When prompted by the Analyzer, remove the RW standard and insert the SDBS standard into the iOS
System, and press Next.
GE Analytical Instruments ©2010
130 of 226
DLM 74001-04 Rev. A