Jp Protocol Worksheet - GE Sievers 500 RL Operation And Maintenance Manual

On-line total organic carbon analyzer
Hide thumbs Also See for Sievers 500 RL:
Table of Contents

Advertisement

Chapter 6: Calibration and System Protocols

JP Protocol Worksheet

Company Name ____________________________
Analyst Name ______________________________
Analyzer Serial Number ______________________
Standards Set Lot No. ___________________________________
TOC Blank Average_____________________________________ (Line 1)
500 ppbC SDBS Average_________________________________ (Line 2)
SDBS Blank-Corrected Average ___________________________ (Line 3)
SDBS Average (Line 2) – TOC Blank Average (Line 1) = SDBS Blank-Corrected Average (Line 3)
Average Temperature-Corrected Conductivity of 29.4 μS/cm __________________ (Line 4)
Temperature __________________________________________________________ (Line 4)
RSD__________________________________________________________________ (Line 5)
Difference ____________________________________________________________ (Line 6)
Acceptance Criteria for JP Protocol TOC (Only):
TOC Blank Average  250 ppb
Blank-Corrected Average  450 ppb
OR
Acceptance Criteria for JP Protocol TOC and Conductivity:
TOC Blank Average  250 ppb
Blank-Corrected Average  450 ppb
Conductivity Difference  ± 
RSD  2%
Sample temperature is between15°C to 30°C

GE Analytical Instruments ©2010
Date _________________________
Firmware Version ______________
Standards Expiration Date _____________
Pass
Fail
Pass
Fail
138 of 226
DLM 74001-04 Rev. A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents